Reflectance Dependent Models – Analog

Philtec Model D21

Technical datasheet for the Philtec Model D21 reflectance-dependent analog fiberoptic displacement sensor. 530 µm spot size, 2.00 mm total operating range, and DC–20 kHz bandwidth.

Philtec analog amplifier Fiberoptic cable and sensor tip
Model D21 amplifier, 914 mm cable, and 0.813 mm probe tip

Technical datasheet for the Philtec Model D21 reflectance-dependent analog fiberoptic displacement sensor. 530 µm spot size, 2.00 mm total operating range, and DC–20 kHz bandwidth.

Spot size530 µm0.021 in target spot
Total range2.00 mm1.75 mm far + 250 µm near
Far side3 mV/µm75 mV/mil nominal
Near side45 mV/µm1.2 mV/µin nominal
BandwidthDC–20 kHz100 Hz to 2 MHz optional

Overview

The Philtec Model D21 is a reflectance-dependent analog fiberoptic sensor for distance, displacement, and vibration on small targets larger than Ø 530 µm. A standard unit includes the amplifier, the sensor tip, and a 914 mm (3 ft) fiberoptic cable. It runs on +12 to +24 VDC and provides a 0 to +5 V analog output.

Output voltage is proportional to both the tip-to-target gap and the reflectivity of the target surface. D models are used where target reflectivity stays constant, such as reciprocating or vibratory motion parallel to the sensor axis.

5.0 V 3.3 V 1.7 V 0.0 V OPTICAL PEAK (5.000 V) NEAR SIDE FAR SIDE Increasing Gap >>>
Figure 1. Near side, optical peak, and far side
  • Near side: steep slope before the peak. High sensitivity (45 mV/µm), limited operating range (250 µm).
  • Optical peak: set to 5.000 V. Zero displacement sensitivity, reflectance dependent; used as the calibration reference.
  • Far side: gradual slope after the peak. Moderate sensitivity (3 mV/µm), greater operating range (1.75 mm).

Specifications

Far side and near side

Parameter Far side (Metric) Far side (Imperial) Near side (Metric) Near side (Imperial)
Total range 1.75 mm 0.070 in 250 µm 10 mils
Linear range 0.70 mm 0.027 in 40 µm 1.6 mils
Nominal standoff 0.75 mm 0.030 in 65 µm 2.5 mils
Nominal sensitivity 3 mV/µm 75 mV/mil 45 mV/µm 1.2 mV/µin
Resolution, DC–200 kHz 1.6 µm 60 µin 0.1 µm 4 µin
Resolution, DC–20 kHz 0.8 µm 30 µin 50 nm 2 µin
Resolution, DC–100 Hz 0.1 µm 4 µin 15 nm 0.6 µin

Tip and cable dimensions

Feature Metric Imperial
Tip outer diameter (Ø C) 0.813 mm 0.032 in
Fiberoptic diameter 0.53 mm 0.021 in
Tip length (C) 38.1 mm 1.5 in
Collar length (B) 12.7 mm 0.5 in
Collar diameter (Ø B) 6.35 mm 0.25 in
Cable length (A) 914 mm 36 in
Cable diameter (Ø A) 4.27 mm 0.168 in
Cable min. bend radius 12.7 mm 0.75 in

Electronics

Light sourceLED, 850 nm
Input voltage+12 to +24 VDC
Input current125 mA max
BandwidthDC–20 kHz standard (3 dB down); up to 2 MHz optional
Isothermal drift0.5%
Amplifier temperature0 to 70 °C
Weight0.7 kg (1.5 lb)

Fiberoptics and probe

Light beam spread30°
Tip material300 series stainless steel
Tip epoxy outgas0.3% @ 200 °C; 2.4% @ 300 °C
Tip pressure15 bar
Tip temperature−55 to 200 °C continuous; to 300 °C intermittent (1–2 hr)
Cable temperature10 to 107 °C
Cable jacketPVC over steel monocoil

Nominal standoff is the gap at the middle of the linear operating range. Published specifications provide nominal values only; actual production units can vary by up to ±15%.

Resolution figures represent best-case performance assuming a flat, smooth, highly reflective target, a perpendicular sensor, the gap of highest sensitivity, and a standard unconnectorized cable.

Response curves

PHILTEC Model D21 — Typical Response (Far Side) 5.0 4.0 3.0 2.0 1.0 0.0 0 0.5 1.0 1.5 2.0 Gap, mm SENSITIVITY = 3 mV/µm
Figure 2. Far side, typical sensitivity 3 mV/µm
  • Linear range, ±1%: 0.4 mm to 1.1 mm.
  • The output peaks at 5.0 V around 0.25 mm, then decreases across the 2.0 mm span.
PHILTEC Model D21 — Typical Response (Near Side) 5.0 4.0 3.0 2.0 1.0 0.0 0 50 100 150 200 250 Gap, micron SENSITIVITY = 45 mV/µm
Figure 3. Near side, typical sensitivity 45 mV/µm
  • Linear range, ±1%: 44.5 µm to 88.9 µm.
  • Output rises steeply from contact and levels near 5.0 V around 250 µm.
Typical Frequency Response — 2 Pole Butterworth Filter 0 -3 -6 -9 100 1000 10000 100000 Frequency, Hz (Gain dB)
Figure 4. Standard 20 kHz 2-pole Butterworth filter stays flat to about 6 kHz
AMPLIFIER DIMENSIONS inches (mm) 5.028 (127.7) 3.24 (82.4) 4 Slots: 0.157 × 0.375 (4 × 9.5 mm) 4.36 (110.8) 1.87 (47.6) 5.48 (139.2)
Figure 5. Standard amplifier enclosure (5.48 × 3.24 × 1.87 in / 139.2 × 82.4 × 47.6 mm)

Distance from the analog output

Inside a linear range, gap change equals the change in millivolts divided by sensitivity (Distance = Δ milliVolts ÷ Sensitivity = µm). Over non-linear ranges, create a lookup table from the XY calibration points or apply a polynomial curve fit.

150.00 µm · 5.906 mils

Questions

What target size and resolution does the Model D21 cover?

The target spot size is 530 µm (0.021 in), designed for targets larger than Ø 530 µm. Best near-side resolution is 15 nm at DC–100 Hz, and 50 nm at the standard DC–20 kHz bandwidth.

When is the near side used instead of the far side?

Use the near side for high sensitivity (45 mV/µm nominal) across a 40 µm linear range (44.5–88.9 µm typical). Use the far side for a wider 0.70 mm linear range (0.4–1.1 mm typical), 3 mV/µm nominal sensitivity, and greater standoff clearance.

How is in-situ calibration performed on different target surfaces?

Adjust the tip-to-target gap until the optical peak output is reached, then use the amplifier gain control to set the peak voltage to full scale (5.000 V). Once set, the factory gap calibration chart applies directly to that target material.

Can the probe tip run hot or under pressure?

The 300-series stainless steel tip is rated from −55 °C to 200 °C continuous (and up to 300 °C intermittent for 1–2 hours) and up to 15 bar pressure. The amplifier electronics are rated 0 to 70 °C.

Staging site at staging.philtec.com. Changes here do not affect the live site.