Reflectance Dependent Models – Analog

Philtec Model D64

Technical datasheet for the Philtec Model D64 reflectance-dependent analog fiberoptic displacement sensor. Ø 1.6 mm spot size, 6 mm total operating range, and DC–20 kHz bandwidth.

Philtec analog amplifier Model D64 fiberoptic cable and sensor tip
Model D64 amplifier, 914 mm cable, and 3.18 mm probe tip

Technical datasheet for the Philtec Model D64 reflectance-dependent analog fiberoptic displacement sensor. Ø 1.6 mm spot size, 6 mm total operating range, and DC–20 kHz bandwidth.

Spot size1.6 mm0.064 in (targets > Ø 1.6 mm)
Total range6.05 mm5.8 mm far + 250 µm near
Far side1.3 mV/µm33 mV/mil nominal
Near side45 mV/µm1.2 mV/µin nominal
BandwidthDC–20 kHz100 Hz to 2 MHz optional

Overview

The Philtec Model D64 is a reflectance-dependent analog fiberoptic sensor for distance, displacement, and vibration of targets larger than Ø 1.6 mm. A standard single-channel unit includes the amplifier and sensor tip with a 914 mm (3 ft) fiberoptic cable. It requires +12 VDC input power (+12 to +24 VDC operating range) and provides a 0 to +5 V analog output with DC–20 kHz bandwidth.

Output voltage is proportional to both the distance between the sensor tip and target and the reflectivity of the target surface. D models are commonly used in applications where target reflectivity stays constant, such as reciprocating or vibratory motion parallel to the axis of the sensor.

5.0 V 3.3 V 1.7 V 0.0 V OPTICAL PEAK (5.000 V) NEAR SIDE FAR SIDE Increasing Gap >>>
Figure 1. Near side, optical peak, and far side
  • Near side: steep slope before the peak. High sensitivity (45 mV/µm nominal; 50 mV/µm feature), limited operating range (250 µm).
  • Optical peak: set to 5.000 V. Zero displacement sensitivity, reflectance dependent; used as the calibration reference.
  • Far side: gradual slope after the peak. Moderate sensitivity (1.3 mV/µm nominal; 1.6 mV/µm feature), greater operating range (5.8 mm).

Specifications

Far side and near side

Parameter Far side (Metric) Far side (Imperial) Near side (Metric) Near side (Imperial)
Total range 5.8 mm 0.230 in 250 µm 0.010 in
Linear range 1.3 mm 0.052 in 38 µm 0.0015 in
Nominal standoff 1.1 mm 0.045 in 53 µm 0.0021 in
Nominal sensitivity 1.3 mV/µm 33 mV/mil 45 mV/µm 1.2 mV/µin
Resolution, DC–200 kHz 1.5 µm 60 µin 45 nm 1.8 µin
Resolution, DC–20 kHz 1 µm 40 µin 30 nm 1.3 µin
Resolution, DC–100 Hz 0.13 µm 5 µin 8 nm 0.3 µin

Tip and cable dimensions

Feature Metric Imperial
Tip outer diameter (Ø C) 3.18 mm 0.125 in
Fiberoptic diameter 1.6 mm 0.066 in
Tip length (C) 76.2 mm 3 in
Collar length (B) 12.7 mm 0.5 in
Collar diameter (Ø B) 6.35 mm 0.25 in
Cable length (A) 914 mm 36 in
Cable diameter (Ø A) 5.7 mm 0.225 in
Cable min. bend radius 19 mm 0.75 in

Electronics

Light sourceLED, 850 nm
Input voltage+12 to +24 VDC
Input current125 mA max
BandwidthDC–20 kHz standard (3 dB down); DC–200 kHz / up to 2 MHz or DC–100 Hz optional
Isothermal drift0.5%
Amplifier temperature0 to 70 °C
Weight0.7 kg (1.5 lb)

Fiberoptics and probe

Light beam spread30°
Tip material300 series stainless steel
Tip epoxy outgas0.3% @ 200 °C; 2.4% @ 300 °C
Tip pressure15 bar
Tip temperature−55 to 200 °C continuous; to 300 °C intermittent (1–2 hr)
Cable temperature10 to 107 °C
Cable jacketPVC over steel monocoil

Nominal standoff is the gap (distance) that places the sensor at the middle of the linear operating range. Standard specifications provide nominal values only; actual production values may vary by as much as ±15%.

Resolution figures represent best-case performance where: the target is flat, smooth, and highly reflective; the sensor is perpendicular to the target; the sensor is gapped to its range of highest sensitivity; and fiberoptic cable lengths are standard with an unconnectorized cable.

Response curves

PHILTEC Model D64 — Typical Response (Far Side) 5.0 4.0 3.0 2.0 1.0 0.0 0 1 2 3 4 5 6 Gap, mm SENSITIVITY = −1.5 mV/µm
Figure 2. Far side, typical sensitivity −1.5 mV/µm (1.3 mV/µm nominal)
  • Linear range, ±1%: 0.4 to 1.8 mm (1.3 mm nominal linear range).
  • Sample front-surface mirror far-side example: sensitivity −6 mV/µm, linear range ±1% = 0.31–0.61 mm, Y-intercept = 6.709 V, noise ripple @ 3.5 VDC (DC–20 kHz) = 6 mV pk-pk.
PHILTEC Model D64 — Typical Response (Near Side) 5.0 4.0 3.0 2.0 1.0 0.0 0 50 100 150 200 250 300 Gap, microns SENSITIVITY = 51 mV/µm
Figure 3. Near side, typical sensitivity 51 mV/µm (45 mV/µm nominal)
  • Linear range, ±1%: 27.9 to 68.6 µm (38 µm nominal linear range).
  • Output rises steeply from contact and levels near 5.0 V toward 250–300 µm.
Typical Frequency Response — 2 Pole Butterworth Filter 0 -3 -6 -9 100 1000 10000 100000 Frequency, Hz (Gain dB)
Figure 4. Standard 20 kHz 2-pole Butterworth filter stays flat to about 6 kHz
AMPLIFIER DIMENSIONS inches (mm) 5.028 (127.7) 3.24 (82.4) 4 Slots: 0.157 × 0.375 (4 × 9.5 mm) 4.36 (110.8) 1.87 (47.6) 5.48 (139.2)
Figure 5. Standard enclosure for D models (5.48 × 3.24 × 1.87 in / 139.2 × 82.4 × 47.6 mm)

Distance from the analog output

A calibration chart is provided with each sensor giving the voltage output response to distance. Inside a linear range, convert the change in voltage output using Distance = Δ milliVolts ÷ Sensitivity = µm. Over non-linear ranges, create a lookup table using the XY calibration data points, or use a polynomial curve fit to map the sensor's output function.

300.00 µm · 11.811 mils

Questions

What target size and resolution does the Model D64 cover?

The Model D64 features a Ø 1.6 mm (0.064 in) target spot size for targets larger than Ø 1.6 mm. Best near-side resolution is 8 nm at DC–100 Hz, 30 nm at DC–20 kHz, and 45 nm at DC–200 kHz. Far-side best-case resolution is 0.13 µm at DC–100 Hz, 1 µm at DC–20 kHz, and 1.5 µm at DC–200 kHz.

When is the near side used instead of the far side?

Use the near side for higher sensitivity (45 mV/µm nominal; 51 mV/µm typical) across a 38 µm linear range (27.9–68.6 µm typical) centered at 53 µm nominal standoff. Use the far side for a wider 1.3 mm linear range (0.4–1.8 mm typical), 1.3 mV/µm nominal sensitivity, and 1.1 mm nominal standoff.

How is in-situ calibration performed on different target surfaces?

Changing target reflectance shifts the voltage output higher or lower. Adjust the sensor's tip-to-target gap until the optical peak output voltage is attained, then use the gain control to set the peak voltage to full scale (5.000 V). After setting the peak to 5 V, the factory gap calibration chart applies for the target being measured.

What factory calibration information is provided with each sensor?

Calibration charts are provided for Near and Far Side regions, typically listing the sensor model and serial number, calibration date, linear sensing range, slope sensitivity, y-intercept of the linear range, and AC noise ripple, with XY calibration data points available upon request.

Can the probe tip run hot or under pressure?

The 300-series stainless steel tip is rated from −55 °C to 200 °C continuous (and up to 300 °C intermittent for 1–2 hours) and up to 15 bar operating pressure. The cable is rated 10 to 107 °C (PVC over steel monocoil), and the amplifier electronics are rated 0 to 70 °C.

Staging site at staging.philtec.com. Changes here do not affect the live site.