Reflectance Dependent Models – Analog

Philtec Model D125

Technical datasheet for the Philtec Model D125 reflectance-dependent analog fiberoptic displacement sensor. Ø 3.18 mm spot size, 15 mm total operating range, and 0.65 mV/µm far side sensitivity.

Philtec analog amplifier Model D125 fiberoptic cable and sensor tip
Model D125 amplifier, 914 mm cable, and 3.96 mm probe tip

Technical datasheet for the Philtec Model D125 reflectance-dependent analog fiberoptic displacement sensor. Ø 3.18 mm spot size, 15 mm total operating range, and 0.65 mV/µm far side sensitivity.

Spot size3.18 mm0.125 in (targets > Ø 3.2 mm)
Total range15 mm14.6 mm (0.575 in) far side
Far side feature0.65 mV/µm6.0 mV/µm (15 mV/mil) spec
Linear range3.3 mm0.130 in far side nominal
BandwidthDC–20 kHzDC–200 kHz / 2 MHz or DC–100 Hz opt.

Overview

The Philtec Model D125 is a reflectance-dependent analog fiberoptic sensor for the measurement of distance, displacement, and vibration of targets larger than Ø 3.2 mm (Ø 3.18 mm / 0.125 inch spot size). Standard single-channel units include the amplifier and sensor tip with a 914 mm (3 foot) fiberoptic cable, require +12 VDC input power (+12 to +24 VDC operating range), and provide a 0 to +5 volt analog output with DC–20 kHz bandwidth.

Output voltage is proportional to both the distance between the sensor tip and target and the reflectivity of the target surface. The Far Side output function starts at the maximum output voltage referred to as the OPTICAL PEAK (set to 5.000 V at factory calibration), and D models are commonly used where target reflectivity stays constant, such as reciprocating or vibratory motion parallel to the axis of the sensor.

5.0 V 3.3 V 1.7 V 0.0 V OPTICAL PEAK (5.000 V) NEAR SIDE FAR SIDE Increasing Gap >>>
Figure 1. Near side, optical peak, and far side
  • Near side: steep slope rising to the optical peak as tip-to-target gap increases from contact.
  • Optical peak: maximum output voltage set to 5.000 V at factory calibration; gain control calibrates other target surfaces to 5 V.
  • Far side: primary calibrated region starting at the optical peak and extending across 14.6 mm (0.575 in) / 15 mm total range.

Specifications

Standard specifications — D125 Far Side (Analog Output 0–5 Volts)

Parameter Metric Imperial
Total range 14.6 mm (15 mm feature) 0.575 in
Linear range* 3.3 mm 0.130 in
Nominal standoff* 2.4 mm 0.095 in
Nominal sensitivity* 6.0 mV/µm (0.65 mV/µm feature) 15 mV/mil
Resolution**, DC–200 kHz 10 µm 400 µin
Resolution**, DC–20 kHz 5 µm 200 µin
Resolution**, DC–100 Hz 0.6 µm 24 µin

Tip and cable dimensions

Feature Metric (mm) Imperial (inch)
Tip outer diameter (Ø C) 3.96 0.156
Fiberoptic diameter 3.18 0.125
Tip length (C) 76.2 3
Collar length (B) 12.7 0.5
Collar diameter (Ø B) 7.11 0.280
Cable length (A) 914 36
Cable diameter (Ø A) 5.89 0.232
Cable min. bend radius 25.4 1

Electronics

Light sourceLED, 850 nm
Input voltage+12 to +24 VDC
Input current125 mA max
BandwidthDC–20 kHz (3 dB down) standard; DC–200 kHz or higher (up to 2 MHz) and DC–100 Hz optional
Isothermal drift0.5%
Operating temperature0 to 70 °C
Weight0.7 kg (1.5 lb)

Fiberoptics

Light beam spread30°
Tip material300 Series SS
Tip epoxy outgas0.3% @ 200 °C; 2.4% @ 300 °C
Tip operating pressure15 bar
Tip operating temperature−55 to 200 °C continuous; to 300 °C intermittent (1–2 hours)
Cable operating temperature10 to 107 °C
Cable jacketPVC over Steel Monocoil

Nominal standoff is the gap (distance) that places the sensor at the middle of the linear operating range. *Standard specifications provide nominal values only; actual production values may vary by as much as ±15%.

**Resolution figures represent best-case performance where: the target is flat, smooth, and highly reflective; the sensor is perpendicular to the target; the sensor is gapped to its range of highest sensitivity; and fiberoptic cable lengths are standard with an unconnectorized cable.

Response curves

PHILTEC Model D125 — Typical Response (Far Side) 5.0 4.0 3.0 2.0 1.0 0.0 0 5 10 15 Gap, mm SENSITIVITY = −0.68 mV/µm
Figure 2. Model D125 typical response — Far Side (−0.68 mV/µm typical)
  • Linear range, Far Side, ±1%: 0.5 – 3.3 mm (3.3 mm nominal linear range).
  • Starts at the 5.000 V optical peak and decreases monotonically across the far side to 15 mm.
PHILTEC Typical Model — Far Side (Front Surface Mirror) 5.0 4.0 3.0 2.0 1.0 0.0 0 0.2 0.4 0.6 0.8 1.0 Gap, mm SENSITIVITY = −6 mV/µm
Figure 3. Typical Far Side calibration example to front surface mirror
  • Linear range, Far Side, ±1%: 0.31 – 0.61 mm; Far Side Y-intercept: 6.709 V.
  • Noise ripple @ 3.5 VDC (DC–20 kHz): 6 mV pk-pk.
Typical Frequency Response — 2 Pole Butterworth Filter 0 -3 -6 -9 100 1000 10000 100000 Frequency, Hz (Gain dB)
Figure 4. Standard 20 kHz 2-pole Butterworth filter stays flat to about 6 kHz
AMPLIFIER DIMENSIONS inches (mm) 5.028 (127.7) 3.24 (82.4) 4 Slots: 0.157 × 0.375 (4 × 9.5 mm) 4.36 (110.8) 1.87 (47.6) 5.48 (139.2)
Figure 5. Standard enclosure for D models (5.48 × 3.24 × 1.87 in / 139.2 × 82.4 × 47.6 mm)

Distance from the analog output

A calibration chart is provided with each sensor giving the voltage output response to distance. There are three ways to derive accurate distance measurements: (a) within the bounds of the linear range, convert the change in voltage output using Distance = Δ milliVolts ÷ Sensitivity = µm; (b) over the non-linear range, create a lookup table using the XY calibration data points; or (c) use a polynomial curve fit to accurately map the sensor's output function.

100.00 µm · 3.937 mils

Factory calibrations

Calibration charts are provided for the Far Side region. A typical factory-supplied calibration chart lists:

  • Sensor model & serial number
  • Date of calibration
  • The linear sensing range
  • The slope sensitivity
  • The y-intercept of the linear range
  • The AC noise ripple

XY calibration data points are available upon request.

End user calibration & dynamic bandwidth

Changing target reflectance shifts the voltage output higher or lower. Factory calibrations set the Peak Voltage to 5.000 V, and a gain control calibrates the output for other target surfaces. Adjust tip-to-target gap to the optical peak voltage, then use the gain control to set full scale (5.000 V), after which the factory gap calibration chart applies.

Analog sensors are fast-responding units for process control and vibration measurements: DC–20 kHz is standard, DC–200 kHz or higher (up to 2 MHz) is optional, and DC–100 Hz (for best resolution) is optional.

Reference conversions: 1 µm = 39.37 µinch · 1 mm = 39.37 mils · 1 mil = 0.001 inch · 1 mil = 25.4 µm · 1 mil = 0.0254 mm.

Contact: PHILTEC, INC., Annapolis, MD USA · Tel: 410-757-4404 · Fax: 410-757-8138 · Email: sensors@philtec.com

Questions

What target size and resolution does the Model D125 cover?

The Model D125 is designed for targets larger than Ø 3.2 mm with a Ø 3.18 mm (0.125 in) target spot size and 15 mm total operating range (14.6 mm / 0.575 in far-side specification). Best-case far-side resolution is 0.6 µm (24 µin) at DC–100 Hz, 5 µm (200 µin) at DC–20 kHz, and 10 µm (400 µin) at DC–200 kHz.

How is in-situ calibration performed on different target surfaces?

Changing target reflectance shifts the voltage output higher or lower. Factory calibrations have the Peak Voltage set to 5.000 V. Adjust the sensor's tip-to-target gap until the optical peak voltage is reached, then use the gain control to set the peak voltage to full scale (5.000 V), after which the factory gap calibration chart applies for that target.

What factory calibration information is provided with each sensor?

Calibration charts are provided for the Far Side region, typically listing the sensor model and serial number, date of calibration, linear sensing range, slope sensitivity, y-intercept of the linear range, and AC noise ripple, with XY calibration data points available upon request.

What bandwidth options are available for dynamic measurements?

DC–20 kHz bandwidth (2-pole Butterworth rolloff, 3 dB down at 20 kHz and flat out to ~6 kHz) is standard. DC–200 kHz or higher (up to 2 MHz) and DC–100 Hz (for best resolution) are optional.

Can the probe tip run hot or under pressure?

The 300-series stainless steel tip is rated from −55 to 200 °C continuous (and up to 300 °C intermittent for 1–2 hours) and up to 15 bar operating pressure, with tip epoxy outgas of 0.3% @ 200 °C and 2.4% @ 300 °C. The cable is rated 10 to 107 °C (PVC over steel monocoil), and the amplifier electronics are rated 0 to 70 °C.