Reflectance Dependent Models – Analog
Philtec Model D171
Technical datasheet for the Philtec Model D171 reflectance-dependent analog fiberoptic displacement sensor. Ø 4.35 mm target spot size, 50 mm total operating range, and 330 mV/mm far side sensitivity.
Technical datasheet for the Philtec Model D171 reflectance-dependent analog fiberoptic displacement sensor. Ø 4.35 mm target spot size, 50 mm total operating range, and 330 mV/mm far side sensitivity.
Overview
The Philtec Model D171 is a reflectance-dependent analog fiberoptic sensor for the measurement of distance, displacement, and vibration of targets larger than Ø 4.4 mm (Ø 4.35 mm / 0.171 inch target spot size, 50 mm total operating range). Standard single-channel units include the amplifier and sensor tip with a 914 mm (3 foot) fiberoptic cable, require +12 VDC input power (+12 to +24 VDC operating range), and provide a 0 to +5 volt analog output with DC–20 kHz bandwidth.
Output voltage is proportional to both the distance between the sensor tip and target and the reflectivity of the target surface. The Far Side output function starts at the maximum output voltage referred to as the OPTICAL PEAK (set to 5.000 V at factory calibration), and incoming light signals are bandpass filtered at 850 ± 22 nm for ambient light rejection. D models are commonly used where target reflectivity stays constant, such as reciprocating or vibratory motion parallel to the axis of the sensor.
- Near side: steep slope rising to the optical peak as tip-to-target gap increases from contact.
- Optical peak: maximum output voltage set to 5.000 V at factory calibration; gain control calibrates other target surfaces to 5 V.
- Far side: primary calibrated region starting at the optical peak and extending across 40.5 mm (1.6 in) / 50 mm total operating range.
Specifications
Standard specifications — D171 Far Side (Analog Output 0–5 Volts)
| Parameter | Metric | Imperial |
|---|---|---|
| Total range | 40.5 mm (50 mm total operating range) | 1.6 in |
| Linear range* | 6.3 mm | 0.25 in |
| Nominal standoff* | 14 mm | 0.55 in |
| Nominal sensitivity* | 330 mV/mm (−0.32 mV/µm typical curve) | 7.4 mV/mil |
| Resolution**, DC–200 kHz | 16 µm | 600 µin |
| Resolution**, DC–20 kHz | 8 µm | 275 µin |
| Resolution**, DC–100 Hz | 2.8 µm | 100 µin |
Tip and cable dimensions
| Feature | Metric (mm) | Imperial (inch) |
|---|---|---|
| Tip outer diameter (Ø C) | 4.75 | 0.187 |
| Fiberoptic diameter | 4.34 | 0.171 |
| Tip length (C) | 76.2 | 3 |
| Collar length (B) | 15.9 | 0.625 |
| Collar diameter (Ø B) | 7.92 | 0.312 |
| Cable length (A) | 914 | 36 |
| Cable diameter (Ø A) | 6.86 | 0.27 |
| Cable min. bend radius | 25.4 | 1 |
Amplifier enclosure dimensions
| Dimension | Metric (mm) | Imperial (inch) |
|---|---|---|
| Body length | 127.7 | 5.028 |
| Flange overall length | 139.2 | 5.48 |
| Body width | 82.4 | 3.24 |
| Mounting slot spacing | 54.6 | 2.15 |
| Height | 47.6 | 1.87 |
| Flange thickness | 1.6 | 0.063 |
| Mounting slots (4 slots) | 4 × 9.5 | 0.157 × 0.375 |
Electronics
| Light source | LED, 850 nm (bandpass filtered at 850 ± 22 nm) |
|---|---|
| Input voltage | +12 to +24 VDC |
| Input current | 125 mA max |
| Bandwidth | DC–20 kHz (3 dB down) standard; DC–200 kHz or higher (up to 2 MHz) and DC–100 Hz optional |
| Isothermal drift | 0.5% |
| Operating temperature | 0 to 70 °C |
| Weight | 0.7 kg (1.5 lb) |
Fiberoptics
| Light beam spread | 25° |
|---|---|
| Tip material | 300 Series SS |
| Tip epoxy outgas | 0.3% @ 200 °C; 2.4% @ 300 °C |
| Tip operating pressure | 15 bar |
| Tip operating temperature | −55 to 200 °C continuous; to 300 °C intermittent (1–2 hours) |
| Cable operating temperature | 10 to 107 °C |
| Cable jacket | PVC over Steel Monocoil |
Nominal standoff is the gap (distance) that places the sensor at the middle of the linear operating range. *Standard specifications provide nominal values only; actual production values may vary by as much as ±15%.
**Resolution figures represent best-case performance where: the target is flat, smooth, and highly reflective; the sensor is perpendicular to the target; the sensor is gapped to its range of highest sensitivity; and fiberoptic cable lengths are standard with an unconnectorized cable.
Response curves
- Linear range, Far Side, ±1%: 10.6 – 16.0 mm (6.3 mm nominal linear range).
- Starts at the 5.000 V optical peak and decreases across the far side out to 50 mm total operating range.
- Linear range, Far Side, ±1%: 0.31 – 0.61 mm; Far Side Y-intercept: 6.709 V.
- Noise ripple @ 3.5 VDC (DC–20 kHz): 6 mV pk-pk.
- Standard D sensor features a 20 kHz 2-pole Butterworth rolloff (−3 dB at 20 kHz).
- Output remains flat out to approximately 6 kHz.
Converting analog output to distance
Factory calibration charts are provided for Near and Far Side regions, listing sensor model & serial number, calibration date, linear sensing range, slope sensitivity, y-intercept, and AC noise ripple (XY data points available upon request).
- Linear range: Distance (µm) = Δ milliVolts ÷ Sensitivity.
- Non-linear range: create a lookup table from the XY calibration data points.
- Curve fit: use a polynomial curve fit to map the sensor output function.
1,000.0 µm (1.000 mm)
Calibration & Conversions
End user calibration
Changing target reflectance shifts the voltage output higher or lower. Factory calibrations set the Optical Peak to 5.000 V, and a gain control calibrates the sensor to other target surfaces:
- Adjust the tip-to-target gap until the peak output voltage is attained.
- Use the gain control to set the peak voltage to full scale (5.000 V).
- Apply the factory gap calibration chart for precision linear motion measurements on the target material.
Unit conversions & contact
| 1 µm | 39.37 µinch |
|---|---|
| 1 mm | 39.37 mils |
| 1 mil | 0.001 inch = 25.4 µm = 0.0254 mm |
| Manufacturer | PHILTEC, Inc., Annapolis, MD USA |
| Phone / Fax / Email | 410-757-4404 · Fax 410-757-8138 · sensors@philtec.com |
Frequently asked questions
What target size and surface motion suit the Model D171?
The D171 has a Ø 4.35 mm (0.171 in) target spot size and is intended for targets larger than Ø 4.4 mm across a 50 mm total operating range. Because output is proportional to both gap and surface reflectivity, D models are best suited where target reflectivity remains constant, such as reciprocating or vibratory motion parallel to the sensor axis.
How is ambient light rejected?
Incoming light signals are bandpass filtered at a wavelength of 850 ± 22 nm to reject unwanted ambient light interference.
What bandwidth options are available?
DC–20 kHz (3 dB down, flat out to ~6 kHz) is standard. Optional bandwidths include DC–200 kHz or higher (up to 2 MHz) for faster dynamics, and DC–100 Hz for best resolution.