Reflectance Dependent Models – Digital

Philtec Model DMS-D20

Technical datasheet for the Philtec Model DMS-D20 digital reflectance-dependent fiberoptic displacement sensor. Ø 510 micron target spot size, 1.27 mm total operating range, RS232 or USB output, and 2.5 nm peak-to-peak resolution.

Philtec DMS digital sensor with the D20 micro probe Model DMS-D20 fiberoptic cable and micro probe tip
Model DMS-D20 digital sensor, 914 mm cable, and Ø 0.81 mm micro probe tip

Technical datasheet for the Philtec Model DMS-D20 digital reflectance-dependent fiberoptic displacement sensor. Ø 510 micron target spot size, 1.27 mm total operating range, RS232 or USB output, and 2.5 nm peak-to-peak resolution.

Spot size510 µm0.020 in (targets > Ø 510 µm)
Total range1.27 mm120 µm Near · 1.15 mm Far
Best resolution2.5 nmNear Side @ 4096 ADC AVG
Clock rate10,416.75 Hz2.5 to 5,208 samples/sec
InterfaceUSB / RS23225 on-board calibration tables

Overview

The Philtec Model DMS-D20 is a reflectance-dependent digital fiberoptic displacement sensor engineered for measuring distance, displacement, and vibration of small targets larger than Ø 510 µm (0.020 inch target spot size) across a 1.27 mm total operating range. Dynamic light signals reflected from target surfaces are converted to linearized distance outputs by comparing incoming signals against gap calibration tables stored on-board the sensor.

The analog input function includes both Near Side (~0–120 µm) and Far Side (~0.2–1.2 mm) regions. Near Side operation delivers the highest resolution with limited operating range, while Far Side operation provides moderate resolution over a much greater operating range. Output is proportional to both tip-to-target distance and target surface reflectivity.

DMS-D20 — Typical Analog Input Function (D Function) 100 80 60 40 20 0 0 0.25 0.5 0.75 1 1.25 NEAR SIDE FAR SIDE Distance To Target, mm
Figure 1. DMS-D20 analog input function (Near Side & Far Side)
  • Near side (~0–120 µm): steepest slope delivering nanometer-level resolution (down to 2.5 nm pk-pk).
  • Far side (~0.2–1.2 mm): wide 1.15 mm linearized range with 8 nm pk-pk resolution at 2.5 S/s.
  • On-board linearization: microprocessor compares incoming signals to stored calibration tables.

Specifications

Standard specifications — Model DMS-D20 (USB or RS232)

Parameter Near Side Far Side
Total range 120 µm 1.15 mm (1.27 mm combined)
Linear range 120 µm 1.15 mm
Optimal gapping region ~0 – 120 microns ~0.2 to 1.2 mm
Reflectance resolution 0.5%
Temperature resolution 0.06 °C

Peak-to-Peak Resolution* vs. ADC Averaging & Sample Rate

ADC AVG Samples / Sec Near Side Pk-Pk Resolution* Far Side Pk-Pk Resolution*
2 5,208 50 nm 200 nm
16 651 40 nm 150 nm
256 41 10 nm 40 nm
4096 2.5 2.5 nm 8 nm

Tip and cable dimensions

Feature Metric (mm) Imperial (inch)
Tip outer diameter (Ø C) 0.81 0.032
Fiberoptic diameter 0.5 (510 µm spot size) 0.020
Tip length (C) 38.1 1.5
Collar length (B) 12.7 0.5
Collar diameter (Ø B) 6.35 0.25
Cable length (A) 914 36
Cable diameter (Ø A) 4.27 0.168
Cable min. bend radius 19 0.75

Electronics

Light source850 nm
Input voltage+12 VDC
Input current500 mA max
Bandwidth5 kHz
Isothermal drift0.05%
Internal clock rate10,416.75 Hz
Weight1.1 kg (2.4 lbs.)

Fiberoptics

Light beam spread66°
FibersGlass
Cable sheathingPVC / Steel Monocoil
Tip epoxy outgas0.3% @ 200 °C; 2.4% @ 300 °C
Tip operating pressure10 bar
Tip operating temperature−55 to 200 °C continuous; to 300 °C intermittent (1–2 hours)

*Pk-Pk Resolution is the smallest measurable displacement limited by the sensor’s self-noise. Specifications represent best-case performance where: the target is flat, smooth, and highly reflective; the sensor is perpendicular to the target; the sensor is gapped to its range of highest sensitivity (~mid-range); and fiberoptic cable lengths are standard with unconnectorized cables.

DMS Control Software includes a data averaging filter from 2 samples (fastest rate) to 4096 samples (highest resolution). Internally, the sensor reads target data continuously at 10,416.75 Hz: Samples/Sec = 10,416.75 / ADC AVG.

Linearized Digital Output & Accuracy

DMS-D20 Digital Output (Linearized Gap Output) 1.25 1.00 0.75 0.50 0.25 0 0 0.25 0.5 0.75 1 1.25 Gap To Target, mm
Figure 2. DMS-D20 linearized digital distance output (0 to 1.27 mm)
  • 1:1 linearized output across both Near Side (0–120 µm) and Far Side (~0.12–1.27 mm) regions.
  • Transition dip around ~0.12 mm marks the Optical Peak boundary between Near and Far Side calibration tables.
Accuracy / Repeatability (Measured at 2.5 samples/sec) 1 0.5 0 -0.5 0 0.25 0.5 0.75 1 1.25 Gap, mm (Error in Microns) Near Side >>> Far Side
Figure 3. Accuracy / repeatability curve (measured at 2.5 samples/sec)
  • Accuracy is a measure of the sensor’s repeatability, measured at 2.5 samples/sec.
  • Deviation remains near 0 µm across both Near Side (0–120 µm) and Far Side (0.2–1.2 mm) active ranges.

Interactive Sample Rate & Resolution Calculator

2.54 Samples/Sec · Near: 2.5 nm · Far: 8 nm

Three Instrument Enclosures to Choose From

RS232 Output

Model mDMS-D20

mDMS-D20 single-channel enclosure with RS232 output

miniDMS single-channel digital sensor with RS-232 serial communication.

mDMS-D20 enclosure dimensions in millimetres
Overall length140 mm
Body L × W × H129 × 60.5 × 32.6 mm
Mounting slots112 × 35.6 mm (4× 9.5 mm)
USB Output

Model muDMS-D20

muDMS-D20 single-channel enclosure with USB output

miniDMS single-channel digital sensor with USB communication.

muDMS-D20 enclosure dimensions in millimetres
Overall length139.2 mm
Body L × W × H127.7 × 82.4 × 47.6 mm
Mounting slots110.8 × 54.6 mm (4× 9.5 mm)
Two Channel · USB & RS232

Model mu2DMS-D20

mu2DMS-D20 two-channel enclosure with USB and RS232

Two-channel digital DMS instrument with both USB and RS232 outputs.

mu2DMS-D20 enclosure dimensions in millimetres
Overall length173 mm
Body L × W × H162 × 95 × 49 mm
Mounting slots145 × 69 mm (4× 9.5 mm)

DMS Setup & Control Software

DMS Setup and Control software showing live distance and reflectance
DMS Setup and Control, version 1.46

On-board Calibration Tables & Freeware

Philtec provides freeware with every digital sensor purchase for controlling sensors, viewing live distance/reflectance data, and saving measurements to files.

  • 25 calibration table slots: stored directly on-board the sensor.
  • Factory pre-loaded tables: (1) Front surface mirror and (2) Diffuse aluminum target.
  • Custom tables: copy/paste calibration data or create and store new target calibration tables.

Software, Firmware & Contact

DMS sensors can be updated remotely from any PC. Download the latest software, firmware, and tutorial videos at www.philtec.com/downloads/firmware-software.

ManufacturerPHILTEC, Inc., Annapolis, MD USA
Phone / Fax / Email410-757-4404 · Fax 410-757-8138 · sensors@philtec.com

Frequently asked questions

What is the difference between Near Side and Far Side operation on the DMS-D20?

Near Side operation (~0–120 µm) uses the steep rising slope of the analog input curve to achieve the highest peak-to-peak resolution (down to 2.5 nm at 2.5 S/s). Far Side operation (~0.2–1.2 mm, 1.15 mm total range) provides moderate resolution (8 nm at 2.5 S/s) over a roughly 10× wider working distance.

How does ADC averaging affect sample rate and resolution?

The sensor continuously samples internally at 10,416.75 Hz. Setting ADC AVG averages between 2 and 4096 internal readings before outputting data to the PC (Samples/Sec = 10,416.75 / ADC AVG), trading sample speed (5,208 S/s down to 2.5 S/s) for ultra-low peak-to-peak noise.

Which model variant should I select for USB vs. RS232?

Select mDMS-D20 for single-channel RS232 communication, muDMS-D20 for single-channel USB communication, or mu2DMS-D20 for two-channel operation with both USB and RS232 outputs.