Reflectance Dependent Models – Digital
Philtec Model DMS-D47
Technical datasheet for the Philtec Model DMS-D47 digital reflectance-dependent fiberoptic displacement sensor. Ø 1.16 mm target spot size, 6 mm total operating range, RS232 or USB output, and 2.5 nm peak-to-peak resolution.
Technical datasheet for the Philtec Model DMS-D47 digital reflectance-dependent fiberoptic displacement sensor. Ø 1.16 mm target spot size, 6 mm total operating range, RS232 or USB output, and 2.5 nm peak-to-peak resolution.
Overview
The Philtec Model DMS-D47 is a reflectance-dependent digital fiberoptic displacement sensor engineered for measuring distance, displacement, and vibration of small targets larger than Ø 1.2 mm (Ø 1.16 mm / 0.047 inch target spot size) across a 6 mm total operating range. Dynamic light signals reflected from target surfaces are converted to linearized distance outputs by comparing incoming signals against gap calibration tables stored on-board the sensor.
The analog input function includes both Near Side (~50–200 µm) and Far Side (~0.5–3 mm) optimal slope regions. Near Side operation delivers the highest resolution with limited operating range, while Far Side operation provides moderate resolution over a much greater operating range. Output is proportional to both tip-to-target distance and target surface reflectivity.
- Near side (~50–200 µm optimal): 300 µm total/linear range delivering nanometer-level resolution (down to 2.5 nm pk-pk).
- Far side (~0.5–3 mm optimal): 5.7 mm total/linear range with 40 nm pk-pk resolution at 2.5 S/s.
- On-board linearization: microprocessor compares incoming signals to stored calibration tables.
Specifications
Standard specifications — Model DMS-D47 (USB or RS232)
| Parameter | Near Side | Far Side |
|---|---|---|
| Total range | 300 µm | 5.7 mm (6 mm combined) |
| Linear range | 300 µm | 5.7 mm |
| Optimal gapping region | ~50 – 200 microns | ~0.5 to 3 mm |
| Reflectance resolution | 0.5% | |
| Temperature resolution | 0.06 °C | |
Peak-to-Peak Resolution* vs. ADC Averaging & Sample Rate
| ADC AVG | Samples / Sec | Near Side Pk-Pk Resolution* | Far Side Pk-Pk Resolution* |
|---|---|---|---|
| 2 | 5,208 | 50 nm | 1,300 nm |
| 16 | 651 | 40 nm | 900 nm |
| 256 | 41 | 10 nm | 250 nm |
| 4096 | 2.5 | 2.5 nm | 40 nm |
Tip and cable dimensions
| Feature | Metric (mm) | Imperial (inch) |
|---|---|---|
| Tip outer diameter (Ø C) | 1.61 | 0.065 |
| Fiberoptic diameter | 1.16 (spot size) | 0.047 |
| Tip length (C) | 38.1 | 1.5 |
| Collar length (B) | 12.7 | 0.5 |
| Collar diameter (Ø B) | 4.75 | 0.187 |
| Cable length (A) | 914 | 36 |
| Cable diameter (Ø A) | 3.34 | 0.135 |
| Cable min. bend radius | 19 | 0.75 |
Electronics
| Light source | 850 nm |
|---|---|
| Input voltage | +12 VDC |
| Input current | 500 mA max |
| Bandwidth | 5 kHz |
| Isothermal drift | 0.05% |
| Internal clock rate | 10,416.75 Hz |
| Weight | 1.1 kg (2.4 lbs) |
Fiberoptics
| Light beam spread | 30° |
|---|---|
| Fibers | Glass |
| Cable sheathing | PVC / Steel Monocoil |
| Tip epoxy outgas | 0.3% @ 200 °C; 2.4% @ 300 °C |
| Tip operating pressure | 10 bar |
| Tip operating temperature | −55 to 200 °C continuous; to 300 °C intermittent (1–2 hours) |
*Pk-Pk Resolution is the smallest measurable displacement limited by the sensor’s self-noise. Specifications represent best-case performance where: the target is flat, smooth, and highly reflective; the sensor is perpendicular to the target; the sensor is gapped to its range of highest sensitivity (~mid-range); and fiberoptic cable lengths are standard with unconnectorized cables.
DMS Control Software includes a data averaging filter from 2 samples (fastest rate) to 4096 samples (highest resolution). Internally, the sensor reads target data continuously at 10,416.75 Hz: Samples/Sec = 10,416.75 / ADC AVG.
Linearized Digital Output & Accuracy
- 1:1 linearized output across both Near Side (0–300 µm) and Far Side (0.3–6.0 mm) regions.
- On-board calibration tables convert reflected light intensity directly into gap distance.
- Accuracy is a measure of the sensor’s repeatability, measured at 2.5 samples/sec.
- Best linearity is achieved on the steepest slopes: Near Side (~50–200 µm) and Far Side (~0.5–3 mm).
Interactive Sample Rate & Resolution Calculator
2.54 Samples/Sec · Near: 2.5 nm · Far: 40 nm
Three Instrument Enclosures to Choose From
Model mDMS-D47
miniDMS single-channel digital sensor with RS-232 serial communication.
| Overall length | 140 mm |
|---|---|
| Body L × W × H | 129 × 60.5 × 32.6 mm |
| Mounting slots | 112 × 35.6 mm (4× 9.5 mm) |
Model muDMS-D47
miniDMS single-channel digital sensor with USB communication.
| Overall length | 139.2 mm |
|---|---|
| Body L × W × H | 127.7 × 82.4 × 47.6 mm |
| Mounting slots | 110.8 × 54.6 mm (4× 9.5 mm) |
Model mu2DMS-D47
Two-channel digital DMS instrument with both USB and RS232 outputs.
| Overall length | 173 mm |
|---|---|
| Body L × W × H | 162 × 95 × 49 mm |
| Mounting slots | 145 × 69 mm (4× 9.5 mm) |
DMS Setup & Control Software
On-board Calibration Tables & Freeware
Philtec provides freeware with every digital sensor purchase for controlling sensors, viewing live distance/reflectance data, and saving measurements to files.
- 25 calibration table slots: stored directly on-board the sensor.
- Factory pre-loaded tables: (1) Front surface mirror and (2) Diffuse aluminum target.
- Custom tables: copy/paste calibration data or create and store new target calibration tables.
Software, Firmware & Contact
DMS sensors can be updated remotely from any PC. Download the latest software, firmware, and tutorial videos at www.philtec.com/downloads/firmware-software.
| Manufacturer | PHILTEC, Inc., Annapolis, MD USA |
|---|---|
| Phone / Fax / Email | 410-757-4404 · Fax 410-757-8138 · sensors@philtec.com |
Frequently asked questions
What is the difference between Near Side and Far Side operation on the DMS-D47?
Near Side operation (300 µm total range, optimal ~50–200 µm) uses the steep rising slope of the analog input curve to achieve the highest peak-to-peak resolution (down to 2.5 nm at 2.5 S/s). Far Side operation (5.7 mm total range, optimal ~0.5–3 mm) provides moderate resolution (40 nm at 2.5 S/s) over a much wider working distance.
How does ADC averaging affect sample rate and resolution?
The sensor continuously samples internally at 10,416.75 Hz. Setting ADC AVG averages between 2 and 4096 internal readings before outputting data to the PC (Samples/Sec = 10,416.75 / ADC AVG), trading sample speed (5,208 S/s down to 2.5 S/s) for ultra-low peak-to-peak noise.
Which model variant should I select for USB vs. RS232?
Select mDMS-D47 for single-channel RS232 communication, muDMS-D47 for single-channel USB communication, or mu2DMS-D47 for two-channel operation with both USB and RS232 outputs.