Reflectance Dependent Models – Digital

Philtec Model DMS-D240

Technical datasheet for the Philtec Model DMS-D240 digital reflectance-dependent fiberoptic displacement sensor. Ø 6.1 mm target spot size for targets > Ø 6.1 mm, 76 mm total operating range, ambient light rejection, RS232 or USB output, and 0.15 µm Near Side / 0.25 µm Far Side peak-to-peak resolution.

Philtec DMS digital sensor with the D240 probe Model DMS-D240 fiberoptic cable and sensor tip
Model DMS-D240 digital sensor, 914 mm cable, and Ø 7.92 mm probe tip

Technical datasheet for the Philtec Model DMS-D240 digital reflectance-dependent fiberoptic displacement sensor. Ø 6.1 mm target spot size for targets > Ø 6.1 mm, 76 mm total operating range, ambient light rejection, RS232 or USB output, and 0.15 µm Near Side / 0.25 µm Far Side peak-to-peak resolution.

Spot size6.1 mm0.240 in (targets > Ø 6.1 mm)
Total range76 mm13 mm Near · 63 mm Far
Best resolution0.15 / 0.25 µmNear / Far @ 4096 ADC AVG
Clock rate10,416.75 Hz2.5 to 5,208 samples/sec
Light rejection850 ± 22 nmBandpass filtered · 25 on-board tables

Overview

The Philtec Model DMS-D240 is a reflectance-dependent digital fiberoptic displacement sensor engineered for measuring distance, displacement, and vibration of targets larger than Ø 6.1 mm (Ø 6.1 mm / 0.240 inch target spot size; fiberoptic diameter 6.1 mm) across a 76 mm total operating range with built-in ambient light rejection (incoming light bandpass filtered at 850 ± 22 nm). Dynamic light signals reflected from target surfaces are converted to linearized distance outputs by comparing incoming signals against gap calibration tables stored on-board the sensor.

The analog input function includes both Near Side (~1–10 mm optimum) and Far Side (~15–50 mm optimum) regions. Near Side operation delivers highest resolution (down to 0.15 µm pk-pk) over a 13 mm total/linear range, while Far Side operation provides moderate resolution (down to 0.25 µm pk-pk) over a 63 mm total/linear range. Output is proportional to both tip-to-target distance and target surface reflectivity.

DMS-D240 — Typical Analog Input (D Function) 100 80 60 40 20 0 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 NEAR SIDE FAR SIDE Distance To Target, mm
Figure 1. DMS-D240 typical analog input (D Function)
  • Near side (~1–10 mm optimum): 13 mm total/linear range delivering resolution down to 0.15 µm pk-pk.
  • Far side (~15–50 mm optimum): 63 mm total/linear range with 0.25 µm pk-pk resolution at 2.5 S/s.
  • Ambient light rejection: bandpass filtered at 850 ± 22 nm.

Specifications

Standard specifications — Model DMS-D240 (USB or RS232)

Parameter Near Side Far Side
Total range 13 mm 63 mm (76 mm combined)
Linear range 13 mm 63 mm
Optimum gapping region ~1 – 10 mm ~15 to 50 mm
Reflectance resolution 0.5%
Temperature resolution 0.06 °C

Peak-to-Peak Resolution* vs. ADC Averaging & Sample Rate

ADC AVG Samples / Sec Near Side Pk-Pk Resolution* Far Side Pk-Pk Resolution*
2 5,208 2 µm 6.5 µm
16 651 1 µm 2.5 µm
256 41 1.25 µm 0.8 µm
4096 2.5 0.15 µm 0.25 µm

Tip and cable dimensions

Feature Metric (mm) Imperial (inch)
Tip outer diameter (Ø C) 7.92 0.312
Fiberoptic diameter 6.1 (spot size: 6.1 mm) 0.240
Tip length (C) 76.2 3.0
Collar length (B) 19.1 0.75
Collar diameter (Ø B) 9.53 0.375
Cable length (A) 914 36
Cable diameter (Ø A) 8.6 0.340
Cable min. bend radius 40 1.6

Electronics

Light source850 nm
Input voltage+12 VDC
Input current500 mA max
Bandwidth5 kHz
Isothermal drift0.05%
Internal clock rate10,416.75 Hz
Weight1.2 kg (2.8 lbs)

Fiberoptics

Light beam spread25°
FibersGlass
Cable sheathingPVC / Steel Monocoil
Tip epoxy outgas0.3% @ 200 °C; 2.4% @ 300 °C
Tip operating pressure15 bar
Tip operating temperature−55 to 200 °C continuous; to 300 °C intermittent (1–2 hours)

*Pk-Pk Resolution is the smallest measurable displacement limited by the sensor’s self-noise. Specifications represent best-case performance where: the target is flat, smooth, and highly reflective; the sensor is perpendicular to the target; the sensor is gapped to its range of highest sensitivity (~mid-range); and fiberoptic cable lengths are standard with unconnectorized cables.

DMS Control Software includes a data averaging filter from 2 samples (fastest rate) to 4096 samples (highest resolution). Internally, the sensor reads target data continuously at 10,416.75 Hz: Samples/Sec = 10,416.75 / ADC AVG.

Linearized Digital Output & Accuracy

DMS-D240 Digital Output 75 50 25 0 0 15 30 45 60 75 Gap To Target, mm
Figure 2. DMS-D240 linearized digital output (0 to 75 mm plotted)
  • 1:1 linearized distance output across Near Side (13 mm) and Far Side (63 mm) regions.
  • On-board calibration tables convert reflected light intensity directly into gap distance.
Accuracy (Measured at 2.5 samples/sec) 10 5 0 -5 -10 0 15 30 45 60 75 mm (Microns)
Figure 3. Accuracy / repeatability curve (measured at 2.5 samples/sec)
  • Accuracy is a measure of the sensor’s repeatability, measured at 2.5 samples/sec.
  • Best performance is achieved on the steepest slopes: Near Side (~1–10 mm) and Far Side (~15–50 mm).

Interactive Sample Rate & Resolution Calculator

2.54 Samples/Sec · Near: 0.15 µm · Far: 0.25 µm

Three Instrument Enclosures to Choose From

RS232 Output

Model mDMS-D240

mDMS-D240 single-channel enclosure with RS232 output

miniDMS single-channel digital sensor with RS-232 serial communication.

mDMS-D240 enclosure dimensions in millimetres
Overall length140 mm
Body L × W × H129 × 60.5 × 32.6 mm
Mounting slots112 × 35.6 mm (4× 9.5 mm)
USB Output

Model muDMS-D240

muDMS-D240 single-channel enclosure with USB output

miniDMS single-channel digital sensor with USB communication.

muDMS-D240 enclosure dimensions in millimetres
Overall length139.2 mm
Body L × W × H127.7 × 82.4 × 47.6 mm
Mounting slots110.8 × 54.6 mm (4× 9.5 mm)
Two Channel · USB & RS232

Model mu2DMS-D240

mu2DMS-D240 two-channel enclosure with USB and RS232

Two-channel digital DMS instrument with both USB and RS232 outputs.

mu2DMS-D240 enclosure dimensions in millimetres
Overall length173 mm
Body L × W × H162 × 95 × 49 mm (62 mm w/ conn.)
Mounting slots145 × 69 mm (4× 9.5 mm)

DMS Setup & Control Software

DMS Setup and Control software showing live distance and reflectance
DMS Setup and Control, version 1.46

On-board Calibration Tables & Freeware

Philtec provides freeware with every digital sensor purchase for controlling sensors, viewing live distance/reflectance data, and saving measurements to files.

  • 25 calibration table slots: stored directly on-board the sensor.
  • Factory pre-loaded tables: (1) Front surface mirror and (2) Diffuse aluminum target.
  • Custom tables: copy/paste calibration data or create and store new target calibration tables.

Software, Firmware & Contact

DMS sensors can be updated remotely from any PC. Download the latest software, firmware, and tutorial videos at www.philtec.com/downloads/firmware-software.

ManufacturerPHILTEC, Inc., Annapolis, MD USA
Phone / Fax / Email410-757-4404 · Fax 410-757-8138 · sensors@philtec.com

Frequently asked questions

What is the difference between Near Side and Far Side operation on the DMS-D240?

Near Side operation (13 mm total/linear range, optimum ~1–10 mm) uses the steep rising slope of the analog input curve to achieve peak-to-peak resolution down to 0.15 µm at 2.5 S/s. Far Side operation (63 mm total/linear range, optimum ~15–50 mm) provides moderate resolution (0.25 µm at 2.5 S/s) over a wider operating range.

How does ADC averaging affect sample rate and resolution?

The sensor continuously samples internally at 10,416.75 Hz. Setting ADC AVG averages between 2 and 4096 internal readings before sending data out to the PC (Samples/Sec = 10,416.75 / ADC AVG), trading sample speed (5,208 S/s down to 2.5 S/s) for ultra-low peak-to-peak noise.

Which model variant should I select for USB vs. RS232?

Select mDMS-D240 for single-channel RS232 communication, muDMS-D240 for single-channel USB communication, or mu2DMS-D240 for two-channel operation with both USB and RS232 outputs.

Staging site at staging.philtec.com. Changes here do not affect the live site.