86_Cryo Fluid Conversions

When a sensor is submersed in a fluid, light rays diverging from the probe tip become more collimated (parallel) because the refractive index of the fluid is closer to the refractive index of the sensor's optical material than air is.

Download PDF

When a sensor is submersed in a fl uid, light rays diverging from the probe tip are more collimated than they would be in air. This increases the operating range of the sensor. The degree of collimation is proportional to the ratio of the refactive index of the fl uid to that of the calibration medium (air or water).

For example, this chart shows the output function of a model RC100 sensor in three diff erent fl uids: air, water and DOT 5 silicone brake fl uid. The fl uids increase the operating range by approximately 30%. To be precise, the sensor must be calibrated in the same medium in which the measurements will be made.

Gap, mm

Volts

Air Water Brake Fluid

PHILTEC Model RC100 Calibration To Front Surface Mirror

Measurements in Cryogenic Fluids

Philtec sensors have been used to monitor speeds and rotor displacements in a variety of pressurized fl uids, such as:

Liquid Hydrogen, -300°F, 1000 psi max. pressure Liquid Oxygen, -297°F, 5000 psi max. pressure Liquid Methane, -297°F, 7500 psi max. pressure

At the factory, sensors are calibrated in air or water at ambient pressure. The XY calibration data is then mathematically corrected for performance in the operating fl uid.

Medium Refractive Index Vacuum 1.00000 Air 1.00004 Liquid Helium 1.026 Liquid Hydrogen 1.0974 Liquid Nitrogen 1.2053 Liquid Oxygen 1.221 Methane (LCH4) 1.286 Water at 35°C 1.3316 Propane (LPG) Kerosene

CRYOGENIC FLUIDS

Liquid fueled rockets burn pressurized cryogenic fl uids via two turbopumps that mix the fuel with an oxidizer to create thrust. They are:

the OTP – Oxygen Turbopump the FTP – Fuel Turbopump

Many companies have used Philtec’s FODS (Fiber Optic Displacement Sensors) for rotor dynamics measurements in the testing phase of these pumps. This table lists the refractive indices of the various fl uids in use at the present time.

Data Conversions with Calibration in Water

LOX and LCH4 are two fl uids with refl ective indices slightly lower than RI of water. Compared to the operating range in water, the range in these fl uids are shortened in the cryo fl uids.

MAPPING EXAMPLE , CALIBRATION IN WATER

mm in H20 mm in LOX* mm in LCH4** Volts 0.000 0.000 0.000 1.016 0.931 0.981 2.032 1.863 1.962 3.047 2.794 2.943 4.063 3.726 3.924 5.080 4.658 4.906 Formula * RI LOX / RI H๜O = 1.221 ÷ 1.332

Formula ** RI LCH4 / RI H๜O = 1.286 ÷ 1.332

CALIBRATIONS AND DATA CONVERSION TO CRYOGENIC FLUIDS

Calibrations Depending upon the refractive index of the cryo fl uid, the sensor will be calibrated in air or water: whichever is closer to the cryo fl uid’s index.

If the medium is liquid hydrogen (1.0974 Index), calibration in air is the best approximation. If the medium is liquid methane (1.286 index), calibration in water is the best approximation.

Data Conversions with Calibration in Air The XY calibration data (x = mm, Y = Volts) are corrected for performance in the cryo fl uid to be used. Using this formula, the operating range is extended in the fl uid.

mm (Cryo) = mm (Air) x RI Cryo Fluid