Explains how submersion in fluids collimates light rays and extends sensor operating range by approximately 30%. Includes refractive index table, calibration requirements per medium, and Philtec experience with cryogenic rocket engine fluids (LH2, LOX, methane).

Download PDF

When a sensor is submersed in a fl uid, light rays diverging from the probe tip are more collimated than they would be in air. This increases the operating range of the sensor. The degree of collimation is proportional to the ratio of the refactive index of the fl uid to that of air.

For example, this chart shows the output function of a model RC100 sensor in three diff erent fl uids: air, water and DOT 5 silicone brake fl uid. The fl uids increase the operating range by approximately 30%. To be precise, the sensor must be calibrated in the same medium in which the measurements will be made.

Gap, mm

Volts

Air Water Brake Fluid

PHILTEC Model RC100 Calibration To Front Surface Mirror

Measurements in Fluids

CRYOGENIC FLUIDS

Liquid fueled rockets burn pressurized cryogenic fl uids to create thrust. Philtec sensors have been used to monitor speeds and rotor displacements in a variety of pressurized cryogenic fl uids, such as:

Liquid Hydrogen, -300°F, 1000 psi max. pressure Liquid Oxygen, -297°F, 5000 psi max. pressure Liquid Methane, -297°F, 7500 psi max. pressure

At the factory, sensors can be calibrated in air or water. The XY calibration data can be mathematically mapped for performance in the cryo fl uid.

Medium Refractive Index Vacuum 1.00000 Air 1.00004 Liquid Helium 1.026 Liquid Hydrogen 1.0974 Liquid Nitrogen 1.2053 Liquid Oxygen 1.221 Methane (LCH4) 1.286 Water at 35° 1.3316 Propane (LPG) Kerosene

If the medium is vacuum, calibration in air is the best approximation If the medium is liquid hydrogen, calibration is air is the best approximation If the medium is methane, calibration in water is the best approximation

Mapping Example

mm in air Volts mm in LOX Volts 1.221*mm in air 1.221*mm in air 1.221*mm in air 1.221*mm in air