Reflectance Dependent Models – Analog

Philtec Model D169

Technical datasheet for the Philtec Model D169 reflectance-dependent analog fiberoptic displacement sensor. Ø 4.29 mm spot size, 20 mm total operating range, and 0.4 mV/µm far side sensitivity.

Philtec analog amplifier Model D169 fiberoptic cable and sensor tip
Model D169 amplifier, 914 mm cable, and 4.75 mm probe tip

Technical datasheet for the Philtec Model D169 reflectance-dependent analog fiberoptic displacement sensor. Ø 4.29 mm spot size, 20 mm total operating range, and 0.4 mV/µm far side sensitivity.

Spot size4.29 mm0.169 in (targets > Ø 4.3 mm)
Total range20 mm19.5 mm (0.770 in) far side
Far side0.4 mV/µm400 mV/mm (10 mV/mil) nominal
Linear range4.0 mm0.160 in far side nominal
BandwidthDC–20 kHzDC–200 kHz / 2 MHz or DC–100 Hz opt.

Overview

The Philtec Model D169 is a reflectance-dependent analog fiberoptic sensor for the measurement of distance, displacement, and vibration of targets larger than Ø 4.3 mm (Ø 4.29 mm / 0.169 inch target spot size). Standard single-channel units include the amplifier and sensor tip with a 914 mm (3 foot) fiberoptic cable, require +12 VDC input power (+12 to +24 VDC operating range), and provide a 0 to +5 volt analog output with DC–20 kHz bandwidth.

Output voltage is proportional to both the distance between the sensor tip and target and the reflectivity of the target surface. The Far Side output function starts at the maximum output voltage referred to as the OPTICAL PEAK (set to 5.000 V at factory calibration), and incoming light signals are bandpass filtered at 850 ± 22 nm for ambient light rejection. D models are commonly used where target reflectivity stays constant, such as reciprocating or vibratory motion parallel to the axis of the sensor.

5.0 V 3.3 V 1.7 V 0.0 V OPTICAL PEAK (5.000 V) NEAR SIDE FAR SIDE Increasing Gap >>>
Figure 1. Near side, optical peak, and far side
  • Near side: steep slope rising to the optical peak as tip-to-target gap increases from contact.
  • Optical peak: maximum output voltage set to 5.000 V at factory calibration; gain control calibrates other target surfaces to 5 V.
  • Far side: primary calibrated region starting at the optical peak and extending across 19.5 mm (0.770 in) / 20 mm total range.

Specifications

Standard specifications — D169 Far Side (Analog Output 0–5 Volts)

Parameter Metric Imperial
Total range 19.5 mm (20 mm feature) 0.770 in
Linear range* 4.0 mm 0.160 in
Nominal standoff* 2.6 mm 0.105 in
Nominal sensitivity* 400 mV/mm (0.4 mV/µm feature) 10 mV/mil
Resolution**, DC–200 kHz 16 µm 620 µin
Resolution**, DC–20 kHz 8 µm 310 µin
Resolution**, DC–100 Hz 0.9 µm 35 µin

Tip and cable dimensions

Feature Metric (mm) Imperial (inch)
Tip outer diameter (Ø C) 4.75 0.187
Fiberoptic diameter 4.29 0.169
Tip length (C) 76.2 3
Collar length (B) 15.9 0.625
Collar diameter (Ø B) 7.92 0.312
Cable length (A) 914 36
Cable diameter (Ø A) 6.86 0.27
Cable min. bend radius 25.4 1

Electronics

Light sourceLED, 850 nm (bandpass filtered at 850 ± 22 nm)
Input voltage+12 to +24 VDC
Input current125 mA max
BandwidthDC–20 kHz (3 dB down) standard; DC–200 kHz or higher (up to 2 MHz) and DC–100 Hz optional
Isothermal drift0.5%
Operating temperature0 to 70 °C
Weight0.7 kg (1.5 lb)

Fiberoptics

Light beam spread30°
Tip material300 Series SS
Tip epoxy outgas0.3% @ 200 °C; 2.4% @ 300 °C
Tip operating pressure15 bar
Tip operating temperature−55 to 200 °C continuous; to 300 °C intermittent (1–2 hours)
Cable operating temperature10 to 107 °C
Cable jacketPVC over Steel Monocoil

Nominal standoff is the gap (distance) that places the sensor at the middle of the linear operating range. *Standard specifications provide nominal values only; actual production values may vary by as much as ±15%.

**Resolution figures represent best-case performance where: the target is flat, smooth, and highly reflective; the sensor is perpendicular to the target; the sensor is gapped to its range of highest sensitivity; and fiberoptic cable lengths are standard with an unconnectorized cable.

Response curves

PHILTEC Model D169 — Typical Response (Far Side) 5.0 4.0 3.0 2.0 1.0 0.0 0 5 10 15 20 Gap, mm SENSITIVITY = −0.42 mV/µm
Figure 2. Model D169 typical response — Far Side (−0.42 mV/µm typical)
  • Linear range, Far Side, ±1%: 1.27 – 5.38 mm (4.0 mm nominal linear range).
  • Starts at the 5.000 V optical peak and decreases monotonically across the far side to ~19.5–20 mm.
PHILTEC Typical Model — Far Side (Front Surface Mirror) 5.0 4.0 3.0 2.0 1.0 0.0 0 0.2 0.4 0.6 0.8 1.0 Gap, mm SENSITIVITY = −6 mV/µm
Figure 3. Typical Far Side calibration example to front surface mirror
  • Linear range, Far Side, ±1%: 0.31 – 0.61 mm; Far Side Y-intercept: 6.709 V.
  • Noise ripple @ 3.5 VDC (DC–20 kHz): 6 mV pk-pk.
Typical Frequency Response — 2 Pole Butterworth Filter 0 -3 -6 -9 100 1000 10000 100000 Frequency, Hz (Gain dB)
Figure 4. Standard 20 kHz 2-pole Butterworth filter stays flat to about 6 kHz
AMPLIFIER DIMENSIONS inches (mm) 5.028 (127.7) 3.24 (82.4) 4 Slots: 0.157 × 0.375 (4 × 9.5 mm) 4.36 (110.8) 1.87 (47.6) 5.48 (139.2)
Figure 5. Standard enclosure for D models (5.48 × 3.24 × 1.87 in / 139.2 × 82.4 × 47.6 mm)

Distance from the analog output

A calibration chart is provided with each sensor giving the voltage output response to distance. There are three ways to derive accurate distance measurements: (a) within the bounds of the linear range, convert the change in voltage output using Distance = Δ milliVolts ÷ Sensitivity = µm; (b) over the non-linear range, create a lookup table using the XY calibration data points; or (c) use a polynomial curve fit to accurately map the sensor's output function.

1000.00 µm · 39.370 mils

Factory calibrations

Calibration charts are provided for the Far Side region. A typical factory-supplied calibration chart lists:

  • Sensor model & serial number
  • Date of calibration
  • The linear sensing range
  • The slope sensitivity
  • The y-intercept of the linear range
  • The AC noise ripple

XY calibration data points are available upon request.

End user calibration & ambient light rejection

Changing target reflectance shifts the voltage output higher or lower. Factory calibrations set the Peak Voltage to 5.000 V, and a gain control calibrates the output for other target surfaces. Adjust tip-to-target gap to the optical peak voltage, then use the gain control to set full scale (5.000 V), after which the factory gap calibration chart applies.

Incoming light signals are bandpass filtered at 850 ± 22 nm for ambient light rejection. Analog sensors are fast-responding units for process control and vibration measurements: DC–20 kHz is standard, DC–200 kHz or higher (up to 2 MHz) is optional, and DC–100 Hz (for best resolution) is optional.

Reference conversions: 1 µm = 39.37 µinch · 1 mm = 39.37 mils · 1 mil = 0.001 inch · 1 mil = 25.4 µm · 1 mil = 0.0254 mm.

Contact: PHILTEC, INC., Annapolis, MD USA · Tel: 410-757-4404 · Fax: 410-757-8138 · Email: sensors@philtec.com

Questions

What target size and resolution does the Model D169 cover?

The Model D169 is designed for targets larger than Ø 4.3 mm with a Ø 4.29 mm (0.169 in) target spot size and 20 mm total operating range (19.5 mm / 0.770 in far-side specification). Best-case far-side resolution is 0.9 µm (35 µin) at DC–100 Hz, 8 µm (310 µin) at DC–20 kHz, and 16 µm (620 µin) at DC–200 kHz.

How does the Model D169 reject ambient light?

Incoming light signals are bandpass filtered at a wavelength of 850 ± 22 nm, helping reject unwanted ambient optical interference outside that band.

How is in-situ calibration performed on different target surfaces?

Changing target reflectance shifts the voltage output higher or lower. Factory calibrations have the Peak Voltage set to 5.000 V. Adjust the sensor's tip-to-target gap until the optical peak voltage is reached, then use the gain control to set the peak voltage to full scale (5.000 V), after which the factory gap calibration chart applies for that target.

What factory calibration information is provided with each sensor?

Calibration charts are provided for the Far Side region, typically listing the sensor model and serial number, date of calibration, linear sensing range, slope sensitivity, y-intercept of the linear range, and AC noise ripple, with XY calibration data points available upon request.

Can the probe tip run hot or under pressure?

The 300-series stainless steel tip is rated from −55 to 200 °C continuous (and up to 300 °C intermittent for 1–2 hours) and up to 15 bar operating pressure, with tip epoxy outgas of 0.3% @ 200 °C and 2.4% @ 300 °C. The cable is rated 10 to 107 °C (PVC over steel monocoil), and the amplifier electronics are rated 0 to 70 °C.

Staging site at staging.philtec.com. Changes here do not affect the live site.