Reflectance Dependent Models – Analog
Philtec Model D169
Technical datasheet for the Philtec Model D169 reflectance-dependent analog fiberoptic displacement sensor. Ø 4.29 mm spot size, 20 mm total operating range, and 0.4 mV/µm far side sensitivity.
Technical datasheet for the Philtec Model D169 reflectance-dependent analog fiberoptic displacement sensor. Ø 4.29 mm spot size, 20 mm total operating range, and 0.4 mV/µm far side sensitivity.
Overview
The Philtec Model D169 is a reflectance-dependent analog fiberoptic sensor for the measurement of distance, displacement, and vibration of targets larger than Ø 4.3 mm (Ø 4.29 mm / 0.169 inch target spot size). Standard single-channel units include the amplifier and sensor tip with a 914 mm (3 foot) fiberoptic cable, require +12 VDC input power (+12 to +24 VDC operating range), and provide a 0 to +5 volt analog output with DC–20 kHz bandwidth.
Output voltage is proportional to both the distance between the sensor tip and target and the reflectivity of the target surface. The Far Side output function starts at the maximum output voltage referred to as the OPTICAL PEAK (set to 5.000 V at factory calibration), and incoming light signals are bandpass filtered at 850 ± 22 nm for ambient light rejection. D models are commonly used where target reflectivity stays constant, such as reciprocating or vibratory motion parallel to the axis of the sensor.
- Near side: steep slope rising to the optical peak as tip-to-target gap increases from contact.
- Optical peak: maximum output voltage set to 5.000 V at factory calibration; gain control calibrates other target surfaces to 5 V.
- Far side: primary calibrated region starting at the optical peak and extending across 19.5 mm (0.770 in) / 20 mm total range.
Specifications
Standard specifications — D169 Far Side (Analog Output 0–5 Volts)
| Parameter | Metric | Imperial |
|---|---|---|
| Total range | 19.5 mm (20 mm feature) | 0.770 in |
| Linear range* | 4.0 mm | 0.160 in |
| Nominal standoff* | 2.6 mm | 0.105 in |
| Nominal sensitivity* | 400 mV/mm (0.4 mV/µm feature) | 10 mV/mil |
| Resolution**, DC–200 kHz | 16 µm | 620 µin |
| Resolution**, DC–20 kHz | 8 µm | 310 µin |
| Resolution**, DC–100 Hz | 0.9 µm | 35 µin |
Tip and cable dimensions
| Feature | Metric (mm) | Imperial (inch) |
|---|---|---|
| Tip outer diameter (Ø C) | 4.75 | 0.187 |
| Fiberoptic diameter | 4.29 | 0.169 |
| Tip length (C) | 76.2 | 3 |
| Collar length (B) | 15.9 | 0.625 |
| Collar diameter (Ø B) | 7.92 | 0.312 |
| Cable length (A) | 914 | 36 |
| Cable diameter (Ø A) | 6.86 | 0.27 |
| Cable min. bend radius | 25.4 | 1 |
Electronics
| Light source | LED, 850 nm (bandpass filtered at 850 ± 22 nm) |
|---|---|
| Input voltage | +12 to +24 VDC |
| Input current | 125 mA max |
| Bandwidth | DC–20 kHz (3 dB down) standard; DC–200 kHz or higher (up to 2 MHz) and DC–100 Hz optional |
| Isothermal drift | 0.5% |
| Operating temperature | 0 to 70 °C |
| Weight | 0.7 kg (1.5 lb) |
Fiberoptics
| Light beam spread | 30° |
|---|---|
| Tip material | 300 Series SS |
| Tip epoxy outgas | 0.3% @ 200 °C; 2.4% @ 300 °C |
| Tip operating pressure | 15 bar |
| Tip operating temperature | −55 to 200 °C continuous; to 300 °C intermittent (1–2 hours) |
| Cable operating temperature | 10 to 107 °C |
| Cable jacket | PVC over Steel Monocoil |
Nominal standoff is the gap (distance) that places the sensor at the middle of the linear operating range. *Standard specifications provide nominal values only; actual production values may vary by as much as ±15%.
**Resolution figures represent best-case performance where: the target is flat, smooth, and highly reflective; the sensor is perpendicular to the target; the sensor is gapped to its range of highest sensitivity; and fiberoptic cable lengths are standard with an unconnectorized cable.
Response curves
- Linear range, Far Side, ±1%: 1.27 – 5.38 mm (4.0 mm nominal linear range).
- Starts at the 5.000 V optical peak and decreases monotonically across the far side to ~19.5–20 mm.
- Linear range, Far Side, ±1%: 0.31 – 0.61 mm; Far Side Y-intercept: 6.709 V.
- Noise ripple @ 3.5 VDC (DC–20 kHz): 6 mV pk-pk.
Distance from the analog output
A calibration chart is provided with each sensor giving the voltage output response to distance. There are three ways to derive accurate distance measurements: (a) within the bounds of the linear range, convert the change in voltage output using Distance = Δ milliVolts ÷ Sensitivity = µm; (b) over the non-linear range, create a lookup table using the XY calibration data points; or (c) use a polynomial curve fit to accurately map the sensor's output function.
Factory calibrations
Calibration charts are provided for the Far Side region. A typical factory-supplied calibration chart lists:
- Sensor model & serial number
- Date of calibration
- The linear sensing range
- The slope sensitivity
- The y-intercept of the linear range
- The AC noise ripple
XY calibration data points are available upon request.
End user calibration & ambient light rejection
Changing target reflectance shifts the voltage output higher or lower. Factory calibrations set the Peak Voltage to 5.000 V, and a gain control calibrates the output for other target surfaces. Adjust tip-to-target gap to the optical peak voltage, then use the gain control to set full scale (5.000 V), after which the factory gap calibration chart applies.
Incoming light signals are bandpass filtered at 850 ± 22 nm for ambient light rejection. Analog sensors are fast-responding units for process control and vibration measurements: DC–20 kHz is standard, DC–200 kHz or higher (up to 2 MHz) is optional, and DC–100 Hz (for best resolution) is optional.
Reference conversions: 1 µm = 39.37 µinch · 1 mm = 39.37 mils · 1 mil = 0.001 inch · 1 mil = 25.4 µm · 1 mil = 0.0254 mm.
Contact: PHILTEC, INC., Annapolis, MD USA · Tel: 410-757-4404 · Fax: 410-757-8138 · Email: sensors@philtec.com
Questions
What target size and resolution does the Model D169 cover?
The Model D169 is designed for targets larger than Ø 4.3 mm with a Ø 4.29 mm (0.169 in) target spot size and 20 mm total operating range (19.5 mm / 0.770 in far-side specification). Best-case far-side resolution is 0.9 µm (35 µin) at DC–100 Hz, 8 µm (310 µin) at DC–20 kHz, and 16 µm (620 µin) at DC–200 kHz.
How does the Model D169 reject ambient light?
Incoming light signals are bandpass filtered at a wavelength of 850 ± 22 nm, helping reject unwanted ambient optical interference outside that band.
How is in-situ calibration performed on different target surfaces?
Changing target reflectance shifts the voltage output higher or lower. Factory calibrations have the Peak Voltage set to 5.000 V. Adjust the sensor's tip-to-target gap until the optical peak voltage is reached, then use the gain control to set the peak voltage to full scale (5.000 V), after which the factory gap calibration chart applies for that target.
What factory calibration information is provided with each sensor?
Calibration charts are provided for the Far Side region, typically listing the sensor model and serial number, date of calibration, linear sensing range, slope sensitivity, y-intercept of the linear range, and AC noise ripple, with XY calibration data points available upon request.
Can the probe tip run hot or under pressure?
The 300-series stainless steel tip is rated from −55 to 200 °C continuous (and up to 300 °C intermittent for 1–2 hours) and up to 15 bar operating pressure, with tip epoxy outgas of 0.3% @ 200 °C and 2.4% @ 300 °C. The cable is rated 10 to 107 °C (PVC over steel monocoil), and the amplifier electronics are rated 0 to 70 °C.