Reflectance Dependent Models – Analog
Philtec Model D47
Technical datasheet for the Philtec Model D47 reflectance-dependent analog fiberoptic displacement sensor. 1.16 mm spot size, 5.0 mm total operating range, and DC–20 kHz bandwidth.
Technical datasheet for the Philtec Model D47 reflectance-dependent analog fiberoptic displacement sensor. 1.16 mm spot size, 5.0 mm total operating range, and DC–20 kHz bandwidth.
Overview
The Philtec Model D47 is a reflectance-dependent analog fiberoptic sensor for distance, displacement, and vibration on small targets larger than Ø 1.2 mm (1.16 mm spot size). A standard unit includes the amplifier, the sensor tip, and a 914 mm (3 ft) fiberoptic cable. It runs on +12 to +24 VDC and provides a 0 to +5 V analog output.
Output voltage is proportional to both the tip-to-target gap and the reflectivity of the target surface. D models are used where target reflectivity stays constant, such as reciprocating or vibratory motion parallel to the sensor axis.
- Near side: steep slope before the peak. High sensitivity (45 mV/µm), limited operating range (300 µm).
- Optical peak: set to 5.000 V. Zero displacement sensitivity, reflectance dependent; used as the calibration reference.
- Far side: gradual slope after the peak. Moderate sensitivity (1.6 mV/µm), greater operating range (4.7 mm).
Specifications
Far side and near side
| Parameter | Far side (Metric) | Far side (Imperial) | Near side (Metric) | Near side (Imperial) |
|---|---|---|---|---|
| Total range | 4.7 mm | 0.175 in | 300 µm | 0.012 in |
| Linear range | 1.3 mm | 0.050 in | 46 µm | 0.0018 in |
| Nominal standoff | 1.1 mm | 0.044 in | 55 µm | 0.0022 in |
| Nominal sensitivity | 1.6 mV/µm | 41 mV/mil | 45 mV/µm | 1.2 mV/µin |
| Resolution, DC–200 kHz | 1.8 µm | 70 µin | 70 nm | 2 µin |
| Resolution, DC–20 kHz | 0.9 µm | 35 µin | 35 nm | 1 µin |
| Resolution, DC–100 Hz | 0.1 µm | 5 µin | 10 nm | 0.4 µin |
Tip and cable dimensions
| Feature | Metric | Imperial |
|---|---|---|
| Tip outer diameter (Ø C) | 1.61 mm | 0.065 in |
| Fiberoptic diameter | 1.16 mm | 0.047 in |
| Tip length (C) | 38.1 mm | 1.5 in |
| Collar length (B) | 12.7 mm | 0.5 in |
| Collar diameter (Ø B) | 4.75 mm | 0.187 in |
| Cable length (A) | 914 mm | 36 in |
| Cable diameter (Ø A) | 3.34 mm | 0.135 in |
| Cable min. bend radius | 19 mm | 0.75 in |
Electronics
| Light source | LED, 850 nm |
|---|---|
| Input voltage | +12 to +24 VDC |
| Input current | 125 mA max |
| Bandwidth | DC–20 kHz standard (3 dB down); up to 2 MHz optional |
| Isothermal drift | 0.5% |
| Amplifier temperature | 0 to 70 °C |
| Weight | 0.7 kg (1.5 lb) |
Fiberoptics and probe
| Light beam spread | 30° |
|---|---|
| Tip material | 300 series stainless steel |
| Tip epoxy outgas | 0.3% @ 200 °C; 2.4% @ 300 °C |
| Tip pressure | 15 bar |
| Tip temperature | −55 to 200 °C continuous; to 300 °C intermittent (1–2 hr) |
| Cable temperature | 10 to 107 °C |
| Cable jacket | PVC over steel monocoil |
Nominal standoff is the gap at the middle of the linear operating range. Published specifications provide nominal values only; actual production units can vary by up to ±15%.
Resolution figures represent best-case performance assuming a flat, smooth, highly reflective target, a perpendicular sensor, the gap of highest sensitivity, and a standard unconnectorized cable.
Response curves
- Linear range, ±1%: 0.4 mm to 2.1 mm.
- The output peaks at 5.0 V around 0.3 mm, then decreases across the 5.0 mm span.
- Nominal linear range: 46 µm (0.0018 in) centered at 55 µm standoff.
- Output rises steeply from contact and levels near 5.0 V around 300 µm.
Distance from the analog output
Inside a linear range, gap change equals the change in millivolts divided by sensitivity (Distance = Δ milliVolts ÷ Sensitivity = µm). Over non-linear ranges, create a lookup table from the XY calibration points or apply a polynomial curve fit.
Questions
What target size and resolution does the Model D47 cover?
The target spot size is 1.16 mm (0.047 in), designed for targets larger than Ø 1.2 mm. Best near-side resolution is 10 nm at DC–100 Hz, and 35 nm at the standard DC–20 kHz bandwidth.
When is the near side used instead of the far side?
Use the near side for high sensitivity (45 mV/µm nominal) across a 46 µm linear range centered at 55 µm standoff. Use the far side for a wide 1.3 mm linear range (0.4–2.1 mm typical), 1.6 mV/µm nominal sensitivity, and greater standoff clearance (1.1 mm nominal).
How is in-situ calibration performed on different target surfaces?
Adjust the tip-to-target gap until the optical peak output is reached, then use the amplifier gain control to set the peak voltage to full scale (5.000 V). Once set, the factory gap calibration chart applies directly to that target material.
Can the probe tip run hot or under pressure?
The 300-series stainless steel tip is rated from −55 °C to 200 °C continuous (and up to 300 °C intermittent for 1–2 hours) and up to 15 bar pressure. The amplifier electronics are rated 0 to 70 °C.