Reflectance Dependent Models – Analog

Philtec Model D63

Technical datasheet for the Philtec Model D63 reflectance-dependent analog fiberoptic displacement sensor. 1.6 mm spot size, 3.0 mm total operating range, and DC–20 kHz bandwidth with high resolution.

Philtec analog amplifier Model D63 fiberoptic cable and sensor tip
Model D63 amplifier, 914 mm cable, and 3.18 mm probe tip

Technical datasheet for the Philtec Model D63 reflectance-dependent analog fiberoptic displacement sensor. 1.6 mm spot size, 3.0 mm total operating range, and DC–20 kHz bandwidth with high resolution.

Spot size1.6 mm0.063 in (targets > Ø 1.6 mm)
Total range3.00 mm2.85 mm far + 150 µm near
Far side3.2 mV/µm80 mV/mil nominal
Near side100 mV/µm2.5 mV/µin nominal
BandwidthDC–20 kHz100 Hz to 2 MHz optional

Overview

The Philtec Model D63 is a reflectance-dependent analog fiberoptic sensor for distance, displacement, and vibration of targets larger than Ø 1.6 mm with high resolution. A standard single-channel unit includes the amplifier, the sensor tip, and a 914 mm (3 ft) fiberoptic cable. It runs on +12 to +24 VDC and provides a 0 to +5 V analog output.

Output voltage is proportional to both the tip-to-target gap and the reflectivity of the target surface. D models are used where target reflectivity stays constant, such as reciprocating or vibratory motion parallel to the sensor axis.

5.0 V 3.3 V 1.7 V 0.0 V OPTICAL PEAK (5.000 V) NEAR SIDE FAR SIDE Increasing Gap >>>
Figure 1. Near side, optical peak, and far side
  • Near side: steep slope before the peak. High sensitivity (100 mV/µm nominal), limited operating range (150 µm).
  • Optical peak: set to 5.000 V. Zero displacement sensitivity, reflectance dependent; used as the calibration reference.
  • Far side: gradual slope after the peak. Moderate sensitivity (3.2 mV/µm nominal), greater operating range (2.85 mm).

Specifications

Far side and near side

Parameter Far side (Metric) Far side (Imperial) Near side (Metric) Near side (Imperial)
Total range 2.85 mm 0.112 in 150 µm 0.006 in
Linear range 0.7 mm 0.028 in 20 µm 0.0008 in
Nominal standoff 0.5 mm 0.021 in 28 µm 0.0011 in
Nominal sensitivity 3.2 mV/µm 80 mV/mil 100 mV/µm 2.5 mV/µin
Resolution, DC–200 kHz 0.6 µm 24 µin 26 nm 1.0 µin
Resolution, DC–20 kHz 0.3 µm 12 µin 13 nm 0.5 µin
Resolution, DC–100 Hz 0.13 µm 5 µin 2 nm 0.08 µin

Tip and cable dimensions

Feature Metric Imperial
Tip outer diameter (Ø C) 3.18 mm 0.125 in
Fiberoptic diameter 1.6 mm 0.063 in
Tip length (C) 76.2 mm 3 in
Collar length (B) 12.7 mm 0.5 in
Collar diameter (Ø B) 6.35 mm 0.25 in
Cable length (A) 914 mm 36 in
Cable diameter (Ø A) 5.7 mm 0.225 in
Cable min. bend radius 19 mm 0.75 in

Electronics

Light sourceLED, 850 nm
Input voltage+12 to +24 VDC
Input current125 mA max
BandwidthDC–20 kHz standard (3 dB down); up to 2 MHz optional
Isothermal drift0.5%
Amplifier temperature0 to 70 °C
Weight0.7 kg (1.5 lb)

Fiberoptics and probe

Light beam spread66°
Tip material300 series stainless steel
Tip epoxy outgas0.3% @ 200 °C; 2.4% @ 300 °C
Tip pressure15 bar
Tip temperature−55 to 200 °C continuous; to 300 °C intermittent (1–2 hr)
Cable temperature10 to 107 °C
Cable jacketPVC over steel monocoil

Nominal standoff is the gap at the middle of the linear operating range. Published specifications provide nominal values only; actual production units can vary by up to ±15%.

Resolution figures represent best-case performance assuming a flat, smooth, highly reflective target, a perpendicular sensor, the gap of highest sensitivity, and a standard unconnectorized cable.

Response curves

PHILTEC Model D63 — Typical Response (Far Side) 5.0 4.0 3.0 2.0 1.0 0.0 0 0.5 1 1.5 2 2.5 3 Gap, mm SENSITIVITY = −2.7 mV/µm
Figure 2. Far side, typical sensitivity −2.7 mV/µm (3.2 mV/µm nominal)
  • Linear range, ±1%: 0.27 mm to 0.92 mm (0.7 mm nominal linear range).
  • Output peaks at 5.0 V around 0.15 mm, then decreases across the 3.0 mm span.
PHILTEC Model D63 — Typical Response (Near Side) 5.0 4.0 3.0 2.0 1.0 0.0 0 25 50 75 100 125 150 Gap, microns SENSITIVITY = 83 mV/µm
Figure 3. Near side, typical sensitivity 83 mV/µm (100 mV/µm nominal)
  • Linear range, ±1%: 17.8 µm to 43.2 µm (20 µm nominal linear range).
  • Output rises steeply from contact and levels near 5.0 V around 150 µm.
Typical Frequency Response — 2 Pole Butterworth Filter 0 -3 -6 -9 100 1000 10000 100000 Frequency, Hz (Gain dB)
Figure 4. Standard 20 kHz 2-pole Butterworth filter stays flat to about 6 kHz
AMPLIFIER DIMENSIONS inches (mm) 5.028 (127.7) 3.24 (82.4) 4 Slots: 0.157 × 0.375 (4 × 9.5 mm) 4.36 (110.8) 1.87 (47.6) 5.48 (139.2)
Figure 5. Standard amplifier enclosure (5.48 × 3.24 × 1.87 in / 139.2 × 82.4 × 47.6 mm)

Distance from the analog output

Inside a linear range, gap change equals the change in millivolts divided by sensitivity (Distance = Δ milliVolts ÷ Sensitivity = µm). Over non-linear ranges, create a lookup table from the XY calibration points or apply a polynomial curve fit.

125.00 µm · 4.921 mils

Questions

What target size and resolution does the Model D63 cover?

The target spot size is 1.6 mm (0.063 in), designed for targets larger than Ø 1.6 mm with high resolution. Best near-side resolution is 2 nm at DC–100 Hz, and 13 nm at the standard DC–20 kHz bandwidth.

When is the near side used instead of the far side?

Use the near side for ultra-high sensitivity (100 mV/µm nominal; 83 mV/µm typical) across a 20 µm linear range (17.8–43.2 µm typical) centered at 28 µm standoff. Use the far side for a wider 0.7 mm linear range (0.27–0.92 mm typical), 3.2 mV/µm nominal sensitivity, and 0.5 mm standoff clearance.

How is in-situ calibration performed on different target surfaces?

Adjust the tip-to-target gap until the optical peak output is reached, then use the amplifier gain control to set the peak voltage to full scale (5.000 V). Once set, the factory gap calibration chart applies directly to that target material.

Can the probe tip run hot or under pressure?

The 300-series stainless steel tip is rated from −55 °C to 200 °C continuous (and up to 300 °C intermittent for 1–2 hours) and up to 15 bar pressure. The amplifier electronics are rated 0 to 70 °C.

Staging site at staging.philtec.com. Changes here do not affect the live site.