Reflectance Dependent Models – Digital

Philtec Model DMS-D125

Technical datasheet for the Philtec Model DMS-D125 digital reflectance-dependent fiberoptic displacement sensor. Ø 3.18 mm target spot size for targets > Ø 3.2 mm, 15 mm total operating range, RS232 or USB output, and 1 nm Near Side / 15 nm Far Side peak-to-peak resolution.

Philtec DMS digital sensor with the D125 probe Model DMS-D125 fiberoptic cable and sensor tip
Model DMS-D125 digital sensor, 914 mm cable, and Ø 3.96 mm probe tip

Technical datasheet for the Philtec Model DMS-D125 digital reflectance-dependent fiberoptic displacement sensor. Ø 3.18 mm target spot size for targets > Ø 3.2 mm, 15 mm total operating range, RS232 or USB output, and 1 nm Near Side / 15 nm Far Side peak-to-peak resolution.

Spot size3.18 mm0.125 in (targets > Ø 3.2 mm)
Total range15 mm400 µm Near · 14.6 mm Far
Best resolution1 nm / 15 nmNear / Far @ 4096 ADC AVG
Clock rate10,416.75 Hz2.5 to 5,208 samples/sec
InterfaceUSB / RS23225 on-board calibration tables

Overview

The Philtec Model DMS-D125 is a reflectance-dependent digital fiberoptic displacement sensor engineered for measuring distance, displacement, and vibration of targets larger than Ø 3.2 mm (Ø 3.18 mm / 0.125 inch target spot size) across a 15 mm total operating range. Dynamic light signals reflected from target surfaces are converted to linearized distance outputs by comparing incoming signals against gap calibration tables stored on-board the sensor.

The analog input function includes both Near Side (~50–300 µm optimum) and Far Side (~0.6–10 mm optimum) regions. Near Side operation delivers highest resolution (down to 1 nm pk-pk) with limited operating range (400 µm), while Far Side operation provides moderate resolution (down to 15 nm pk-pk) over a much greater operating range (14.6 mm). Output is proportional to both tip-to-target distance and target surface reflectivity.

DMS-D125 — Typical Analog Input (D Function) 100 80 60 40 20 0 0 2.5 5 7.5 10 12.5 15 NEAR SIDE FAR SIDE Distance To Target, mm
Figure 1. DMS-D125 analog input function (Near Side & Far Side)
  • Near side (~50–300 µm optimum): 400 µm total/linear range delivering resolution down to 1 nm pk-pk.
  • Far side (~0.6–10 mm optimum): 14.6 mm total/linear range with 15 nm pk-pk resolution at 2.5 S/s.
  • On-board linearization: microprocessor compares incoming signals to stored calibration tables.

Specifications

Standard specifications — Model DMS-D125 (USB or RS232)

Parameter Near Side Far Side
Total range 400 µm 14.6 mm (15 mm combined)
Linear range 400 µm 14.6 mm
Optimum gapping region ~50 – 300 microns ~0.6 to 10 mm
Reflectance resolution 0.5%
Temperature resolution 0.06 °C

Peak-to-Peak Resolution* vs. ADC Averaging & Sample Rate

ADC AVG Samples / Sec Near Side Pk-Pk Resolution* Far Side Pk-Pk Resolution*
2 5,208 50 nm 750 nm
16 651 40 nm 500 nm
256 41 10 nm 130 nm
4096 2.5 1 nm 15 nm

Tip and cable dimensions

Feature Metric (mm) Imperial (inch)
Tip outer diameter (Ø C) 3.96 0.156
Fiberoptic diameter 3.18 (spot size) 0.125
Tip length (C) 76.2 3
Collar length (B) 12.7 0.5
Collar diameter (Ø B) 7.11 0.280
Cable length (A) 914 36
Cable diameter (Ø A) 5.89 0.232
Cable min. bend radius 25.4 1

Electronics

Light source850 nm
Input voltage+12 VDC
Input current500 mA max
Bandwidth5 kHz
Isothermal drift0.05%
Internal clock rate10,416.75 Hz
Weight1.1 kg (2.4 lbs)

Fiberoptics

Light beam spread30°
FibersGlass
Cable sheathingPVC / Steel Monocoil
Tip epoxy outgas0.3% @ 200 °C; 2.4% @ 300 °C
Tip operating pressure10 bar
Tip operating temperature−55 to 200 °C continuous; to 300 °C intermittent (1–2 hours)

*Pk-Pk Resolution is the smallest measurable displacement limited by the sensor’s self-noise. Specifications represent best-case performance where: the target is flat, smooth, and highly reflective; the sensor is perpendicular to the target; the sensor is gapped to its range of highest sensitivity (~mid-range); and fiberoptic cable lengths are standard with unconnectorized cables.

DMS Control Software includes a data averaging filter from 2 samples (fastest rate) to 4096 samples (highest resolution). Internally, the sensor reads target data continuously at 10,416.75 Hz: Samples/Sec = 10,416.75 / ADC AVG.

Linearized Digital Output & Accuracy

DMS-D125 Digital Output (Linearized Gap Output) 15 10 5 0 0 5 10 15 Gap To Target, mm
Figure 2. DMS-D125 linearized digital distance output (0 to 15 mm)
  • 1:1 linearized output across both Near Side (0–400 µm) and Far Side (0.4–15.0 mm) regions.
  • On-board calibration tables convert reflected light intensity directly into gap distance.
Accuracy / Repeatability (Measured at 2.5 samples/sec) 2 1 0 -1 -2 0 2 4 6 8 10 12 14 mm (Error in Microns) >>> Far Side
Figure 3. Accuracy / repeatability curve (measured at 2.5 samples/sec)
  • Accuracy is a measure of the sensor’s repeatability, measured at 2.5 samples/sec.
  • Best linearity is achieved on the steepest slopes: Near Side (~50–300 µm) and Far Side (~0.6–10 mm).

Interactive Sample Rate & Resolution Calculator

2.54 Samples/Sec · Near: 1 nm · Far: 15 nm

Three Instrument Enclosures to Choose From

RS232 Output

Model mDMS-D125

mDMS-D125 single-channel enclosure with RS232 output

miniDMS single-channel digital sensor with RS-232 serial communication.

mDMS-D125 enclosure dimensions in millimetres
Overall length140 mm
Body L × W × H129 × 60.5 × 32.6 mm
Mounting slots112 × 35.6 mm (4× 9.5 mm)
USB Output

Model muDMS-D125

muDMS-D125 single-channel enclosure with USB output

miniDMS single-channel digital sensor with USB communication.

muDMS-D125 enclosure dimensions in millimetres
Overall length139.2 mm
Body L × W × H127.7 × 82.4 × 47.6 mm
Mounting slots110.8 × 54.6 mm (4× 9.5 mm)
Two Channel · USB & RS232

Model mu2DMS-D125

mu2DMS-D125 two-channel enclosure with USB and RS232

Two-channel digital DMS instrument with both USB and RS232 outputs.

mu2DMS-D125 enclosure dimensions in millimetres
Overall length173 mm
Body L × W × H162 × 95 × 49 mm (62 mm w/ conn.)
Mounting slots145 × 69 mm (4× 9.5 mm)

DMS Setup & Control Software

DMS Setup and Control software showing live distance and reflectance
DMS Setup and Control, version 1.46

On-board Calibration Tables & Freeware

Philtec provides freeware with every digital sensor purchase for controlling sensors, viewing live distance/reflectance data, and saving measurements to files.

  • 25 calibration table slots: stored directly on-board the sensor.
  • Factory pre-loaded tables: (1) Front surface mirror and (2) Diffuse aluminum target.
  • Custom tables: copy/paste calibration data or create and store new target calibration tables.

Software, Firmware & Contact

DMS sensors can be updated remotely from any PC. Download the latest software, firmware, and tutorial videos at www.philtec.com/downloads/firmware-software.

ManufacturerPHILTEC, Inc., Annapolis, MD USA
Phone / Fax / Email410-757-4404 · Fax 410-757-8138 · sensors@philtec.com

Frequently asked questions

What is the difference between Near Side and Far Side operation on the DMS-D125?

Near Side operation (400 µm total range, optimum ~50–300 µm) uses the steep rising slope of the analog input curve to achieve peak-to-peak resolution down to 1 nm at 2.5 S/s. Far Side operation (14.6 mm total range, optimum ~0.6–10 mm) provides moderate resolution (15 nm at 2.5 S/s) over a much wider working distance.

How does ADC averaging affect sample rate and resolution?

The sensor continuously samples internally at 10,416.75 Hz. Setting ADC AVG averages between 2 and 4096 internal readings before outputting data to the PC (Samples/Sec = 10,416.75 / ADC AVG), trading sample speed (5,208 S/s down to 2.5 S/s) for ultra-low peak-to-peak noise.

Which model variant should I select for USB vs. RS232?

Select mDMS-D125 for single-channel RS232 communication, muDMS-D125 for single-channel USB communication, or mu2DMS-D125 for two-channel operation with both USB and RS232 outputs.

Staging site at staging.philtec.com. Changes here do not affect the live site.