Reflectance Dependent Models – Digital

Philtec Model DMS-D64

Technical datasheet for the Philtec Model DMS-D64 digital reflectance-dependent fiberoptic displacement sensor. Ø 1.6 mm target spot size, 6 mm total operating range, RS232 or USB output, and 0.3 nm peak-to-peak resolution.

Philtec DMS digital sensor with the D64 probe Model DMS-D64 fiberoptic cable and sensor tip
Model DMS-D64 digital sensor, 914 mm cable, and Ø 3.18 mm probe tip

Technical datasheet for the Philtec Model DMS-D64 digital reflectance-dependent fiberoptic displacement sensor. Ø 1.6 mm target spot size, 6 mm total operating range, RS232 or USB output, and 0.3 nm peak-to-peak resolution.

Spot size1.6 mm0.064 in (targets > Ø 1.6 mm)
Total range6 mm400 µm Near · 5.6 mm Far
Best resolution0.3 nmNear Side @ 4096 ADC AVG
Clock rate10,416.75 Hz2.5 to 5,208 samples/sec
InterfaceUSB / RS23225 on-board calibration tables

Overview

The Philtec Model DMS-D64 is a reflectance-dependent digital fiberoptic displacement sensor engineered for measuring distance, displacement, and vibration of small targets larger than Ø 1.6 mm (Ø 1.6 mm / 0.064 inch target spot size) across a 6 mm total operating range. Dynamic light signals reflected from target surfaces are converted to linearized distance outputs by comparing incoming signals against gap calibration tables stored on-board the sensor.

The analog input function includes both Near Side (~50–300 µm) and Far Side (~0.5–4 mm) optimal slope regions. Near Side operation delivers highest resolution (down to 0.3 nm pk-pk) with limited operating range, while Far Side operation provides moderate resolution over a much greater operating range. Output is proportional to both tip-to-target distance and target surface reflectivity.

DMS-D64 — Typical Analog Input (D Function) 100 80 60 40 20 0 0 1 2 3 4 5 6 NEAR SIDE FAR SIDE Distance To Target, mm
Figure 1. DMS-D64 analog input function (Near Side & Far Side)
  • Near side (~50–300 µm optimal): 400 µm total/linear range delivering resolution down to 0.3 nm pk-pk.
  • Far side (~0.5–4 mm optimal): 5.6 mm total/linear range with 15 nm pk-pk resolution at 2.5 S/s.
  • On-board linearization: microprocessor compares incoming signals to stored calibration tables.

Specifications

Standard specifications — Model DMS-D64 (USB or RS232)

Parameter Near Side Far Side
Total range 400 µm 5.6 mm (6 mm combined)
Linear range 400 µm 5.6 mm
Optimal gapping region ~50 – 300 microns ~0.5 to 4 mm
Reflectance resolution 0.5%
Temperature resolution 0.06 °C

Peak-to-Peak Resolution* vs. ADC Averaging & Sample Rate

ADC AVG Samples / Sec Near Side Pk-Pk Resolution* Far Side Pk-Pk Resolution*
2 5,208 10 nm 400 nm
16 651 3 nm 150 nm
256 41 1 nm 50 nm
4096 2.5 0.3 nm 15 nm

Tip and cable dimensions

Feature Metric (mm) Imperial (inch)
Tip outer diameter (Ø C) 3.18 0.125
Fiberoptic diameter 1.6 (spot size) 0.064
Tip length (C) 76.2 3
Collar length (B) 12.7 0.5
Collar diameter (Ø B) 6.35 0.25
Cable length (A) 914 36
Cable diameter (Ø A) 5.7 0.225
Cable min. bend radius 19 0.75

Electronics

Light source850 nm
Input voltage+12 VDC
Input current500 mA max
Bandwidth5 kHz
Isothermal drift0.05%
Internal clock rate10,416.75 Hz
Weight1.1 kg (2.4 lbs)

Fiberoptics

Light beam spread30°
FibersGlass
Cable sheathingPVC / Steel Monocoil
Tip epoxy outgas0.3% @ 200 °C; 2.4% @ 300 °C
Tip operating pressure10 bar
Tip operating temperature−55 to 200 °C continuous; to 300 °C intermittent (1–2 hours)

*Pk-Pk Resolution is the smallest measurable displacement limited by the sensor’s self-noise. Specifications represent best-case performance where: the target is flat, smooth, and highly reflective; the sensor is perpendicular to the target; the sensor is gapped to its range of highest sensitivity (~mid-range); and fiberoptic cable lengths are standard with unconnectorized cables.

DMS Control Software includes a data averaging filter from 2 samples (fastest rate) to 4096 samples (highest resolution). Internally, the sensor reads target data continuously at 10,416.75 Hz: Samples/Sec = 10,416.75 / ADC AVG.

Linearized Digital Output & Accuracy

DMS-D64 Digital Output (Linearized Gap Output) 6 5 4 3 2 1 0 0 1 2 3 4 5 6 Gap To Target, mm
Figure 2. DMS-D64 linearized digital distance output (0 to 6 mm)
  • 1:1 linearized output across both Near Side (0–400 µm) and Far Side (0.4–6.0 mm) regions.
  • On-board calibration tables convert reflected light intensity directly into gap distance.
Accuracy / Repeatability (Measured at 2.5 samples/sec) 1 0.5 0 -0.5 0 1 2 3 4 5 6 Gap, mm (Error in Microns) >>> Far Side
Figure 3. Accuracy / repeatability curve (measured at 2.5 samples/sec)
  • Accuracy is a measure of the sensor’s repeatability, measured at 2.5 samples/sec.
  • Best linearity is achieved on the steepest slopes: Near Side (~50–300 µm) and Far Side (~0.5–4 mm).

Interactive Sample Rate & Resolution Calculator

2.54 Samples/Sec · Near: 0.3 nm · Far: 15 nm

Three Instrument Enclosures to Choose From

RS232 Output

Model mDMS-D64

mDMS-D64 single-channel enclosure with RS232 output

miniDMS single-channel digital sensor with RS-232 serial communication.

mDMS-D64 enclosure dimensions in millimetres
Overall length140 mm
Body L × W × H129 × 60.5 × 32.6 mm
Mounting slots112 × 35.6 mm (4× 9.5 mm)
USB Output

Model muDMS-D64

muDMS-D64 single-channel enclosure with USB output

miniDMS single-channel digital sensor with USB communication.

muDMS-D64 enclosure dimensions in millimetres
Overall length139.2 mm
Body L × W × H127.7 × 82.4 × 47.6 mm
Mounting slots110.8 × 54.6 mm (4× 9.5 mm)
Two Channel · USB & RS232

Model mu2DMS-D64

mu2DMS-D64 two-channel enclosure with USB and RS232

Two-channel digital DMS instrument with both USB and RS232 outputs.

mu2DMS-D64 enclosure dimensions in millimetres
Overall length173 mm
Body L × W × H162 × 95 × 49 mm
Mounting slots145 × 69 mm (4× 9.5 mm)

DMS Setup & Control Software

DMS Setup and Control software showing live distance and reflectance
DMS Setup and Control, version 1.46

On-board Calibration Tables & Freeware

Philtec provides freeware with every digital sensor purchase for controlling sensors, viewing live distance/reflectance data, and saving measurements to files.

  • 25 calibration table slots: stored directly on-board the sensor.
  • Factory pre-loaded tables: (1) Front surface mirror and (2) Diffuse aluminum target.
  • Custom tables: copy/paste calibration data or create and store new target calibration tables.

Software, Firmware & Contact

DMS sensors can be updated remotely from any PC. Download the latest software, firmware, and tutorial videos at www.philtec.com/downloads/firmware-software.

ManufacturerPHILTEC, Inc., Annapolis, MD USA
Phone / Fax / Email410-757-4404 · Fax 410-757-8138 · sensors@philtec.com

Frequently asked questions

What is the difference between Near Side and Far Side operation on the DMS-D64?

Near Side operation (400 µm total range, optimal ~50–300 µm) uses the steep rising slope of the analog input curve to achieve peak-to-peak resolution down to 0.3 nm at 2.5 S/s. Far Side operation (5.6 mm total range, optimal ~0.5–4 mm) provides moderate resolution (15 nm at 2.5 S/s) over a much wider working distance.

How does ADC averaging affect sample rate and resolution?

The sensor continuously samples internally at 10,416.75 Hz. Setting ADC AVG averages between 2 and 4096 internal readings before outputting data to the PC (Samples/Sec = 10,416.75 / ADC AVG), trading sample speed (5,208 S/s down to 2.5 S/s) for ultra-low peak-to-peak noise.

Which model variant should I select for USB vs. RS232?

Select mDMS-D64 for single-channel RS232 communication, muDMS-D64 for single-channel USB communication, or mu2DMS-D64 for two-channel operation with both USB and RS232 outputs.

Staging site at staging.philtec.com. Changes here do not affect the live site.