Reflectance Compensated Models – Analog

Philtec Model RC20

Technical datasheet for the Philtec Model RC20 reflectance-compensated analog fiberoptic displacement sensor. Ø 0.5 mm spot size for targets > Ø 0.5 mm, 1.65 mm total operating range, 4.0 mV/µm (100 mV/mil) nominal sensitivity, 0 to +5 V output, DC–20 kHz standard bandwidth, and resolution down to 0.3 µm.

Philtec analog RC amplifier Model RC20 fiberoptic cable, collar, and fine probe tip
Model RC20 analog amplifier, 914 mm cable, and fine Ø 0.81 mm probe tip

Technical datasheet for the Philtec Model RC20 reflectance-compensated analog fiberoptic displacement sensor. Ø 0.5 mm spot size for targets > Ø 0.5 mm, 1.65 mm total operating range, 4.0 mV/µm (100 mV/mil) nominal sensitivity, 0 to +5 V output, DC–20 kHz standard bandwidth, and resolution down to 0.3 µm.

Spot size0.5 mm0.020 in (targets > Ø 0.5 mm)
Total range1.65 mm0.065 in · Linear: 0.6 mm
Nominal sensitivity4.0 mV/µm100 mV/mil (0–5 V output)
BandwidthDC – 20 kHz3 dB down (2-pole Butterworth)
Best resolution0.3 µm12 µin @ DC–100 Hz

Overview & Reflectance Compensation

The Philtec Model RC20 is a reflectance-compensated analog fiberoptic displacement sensor designed for measuring distance, displacement, and vibration of small targets larger than Ø 0.5 mm. RC Model sensors feature a pair of adjacent fiberoptic detectors in the sensor tip; light reflected off the target travels along two separate paths back to the electronics, where a ratiometric calculation yields a distance measurement independent of varying surface reflectance.

Analog sensors are fast-responding instruments suited to dynamic process control and vibration measurements. Standard single-channel units include the amplifier and sensor tip with a 914 mm (3 ft) glass fiberoptic cable, require +12 VDC input power (+12 to +24 VDC rated), and provide a 0 to +5 V analog output with DC–20 kHz bandwidth (optional DC–200 kHz or higher up to 350 kHz; optional DC–100 Hz for highest resolution).

Typical RC Model Response (0 to +5 Volts) 5 4 3 2 1 0 0 10 20 30 40 50 Increasing Gap
Figure 1. Typical RC model voltage response vs. increasing gap
  • Ratiometric compensation: dual adjacent detectors cancel surface reflectance variations.
  • Nominal standoff: 1.0 mm (0.035 in.) places the sensor at the center of the 0.6 mm (0.025 in.) linear range.
  • Beam spread: 30° light beam spread from an 850 nm LED source.

Specifications

Standard specifications — Model RC20 (0–5 Volts Analog Output)

Parameter Metric Imperial
Total range 1.65 mm 0.065 in.
Linear range* 0.6 mm 0.025 in.
Nominal standoff* 1.0 mm 0.035 in.
Nominal sensitivity* 4.0 mV/µm 100 mV/mil
Resolution** (DC – 200 kHz) 2 µm 80 µin
Resolution** (DC – 20 kHz) 1 µm 40 µin
Resolution** (DC – 100 Hz) 0.3 µm 12 µin
Weight 0.7 kg 1.5 lbs.

Electronics

Light sourceLED, 850 nm
Input voltage+12 to +24 VDC
Input current125 mA max
Bandwidth (3 dB down)DC – 20 kHz (optional 200+ kHz up to 350 kHz; or DC – 100 Hz)
Isothermal drift0.5%
Operating temperature0 to 70 °C

Fiberoptics

Light beam spread30°
Tip material300 Series SS
Tip epoxy outgas0.3% @ 200 °C; 2.4% @ 300 °C
Tip operating pressure15 bar
Tip operating temperature−55 to 200 °C continuous; to 300 °C intermittent (1–2 hours)
Cable jacketPVC over Steel Monocoil (10 to 107 °C)

NOTE: Nominal Standoff = the gap distance that places the sensor at the middle of the linear operating range. *Standard Specifications provide nominal values only; actual production values may vary by as much as ±15%.

**Resolution represents best-case performance where: the target is flat, smooth, and highly reflective; the sensor is perpendicular to the target; the sensor is gapped to its range of highest sensitivity; and fiberoptic cable lengths are standard with unconnectorized cable.

Tip, Cable & Amplifier Dimensions

Feature mm inch
Tip outer diameter, Ø C 0.81 0.032
Fiberoptic Ø (Spot size: 0.5 mm) 0.5 0.020
Tip length, C 38 1.5
Collar length, B 12.7 0.5
Collar diameter, Ø B 6.35 0.25
Cable length, A 914 36
Cable diameter, Ø A 5.2 0.205
Cable min. bend radius 12.7 0.5

Sensor Tip Alignment & Amplifier Enclosure

RC sensors have side-by-side adjacent fiberoptic bundles in the tip face. An alignment flat on the casing is ground parallel to the split between the adjacent fiberoptic bundles and located on the light-transmitting side.

Analog RC amplifier enclosure dimensions in inches and millimeters Alignment flat on the RC sensor tip, parallel to the fiber bundle split
Amplifier body / flange L127.7 / 139.2 mm5.028 / 5.48 in.
Amplifier width / height82.4 / 47.6 mm3.24 / 1.87 in.
Mounting slots (4 places)54.6 × 110.8 mm (4 × 9.5 slots)2.15 × 4.36 in. (0.157 × 0.375 slots)
Flange thickness1.6 mm0.063 in.

Typical Mirror Calibration & Frequency Response

PHILTEC Model RC20 — Serial No. 3082 (Front Surface Mirror) 5.0 4.0 3.0 2.0 1.0 0.0 0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 SENSITIVITY = 3.912 mV/µm Gap, mm
Figure 2. Model RC20 (S/N 3082) calibration to front surface mirror (2/28/2018)
  • Calibration values (S/N 3082): Sensitivity = 3.912 mV/µm; Linear Range ±1% = 0.66–1.30 mm.
  • Y-intercept: −1.051 Volts; Noise ripple @ 2 VDC (DC–20 kHz): 3 mV pk-pk.
Typical Frequency Response — 2-Pole Butterworth Filter 0 -3 -6 -9 100 1000 10000 100000 Frequency, Hz
Figure 3. 2-pole Butterworth filter response (3 dB down @ 20 kHz)
  • With 3 dB point at 20 kHz, output is flat out to approximately 6 kHz.
  • Optional amplifiers set 3 dB down at 200 kHz or higher (up to 350 kHz) or at 100 Hz.

Converting Analog Output to Distance

Each sensor ships with a calibration chart listing model & serial number, calibration date, linear sensing range, slope sensitivity, y-intercept of the linear range, and AC noise ripple (XY calibration points available upon request). Three methods derive distance: (a) within the linear range, Distance (µm) = Δ milliVolts ÷ Sensitivity; (b) over the non-linear range, use a lookup table of XY calibration points; or (c) apply a polynomial curve fit.

501.02 µm (0.5010 mm · 19.73 mil)

Manufacturer & Contact

ManufacturerPHILTEC, Inc., Annapolis, MD USA
Phone / Fax / Email410-757-4404 · Fax 410-757-8138 · sensors@philtec.com

Frequently asked questions

How does reflectance compensation work on the RC20?

The RC20 tip contains a pair of adjacent fiberoptic detectors. Light reflected from the target follows two separate paths back to the electronics, where a ratiometric calculation compensates for variations in target surface reflectance.

What is the purpose of the alignment flat on the RC20 tip/casing?

Because RC sensors use side-by-side adjacent fiberoptic bundles, an alignment flat is ground parallel to the split between the adjacent bundles on the light-transmitting side to aid sensor orientation.

What bandwidth options are available for the RC20 analog amplifier?

DC–20 kHz (2-pole Butterworth rolloff, flat to ~6 kHz) is standard, with optional high-frequency amplifiers at 200 kHz or higher (up to 350 kHz) and low-frequency amplifiers at 100 Hz for highest resolution (0.3 µm / 12 µin).

Staging site at staging.philtec.com. Changes here do not affect the live site.